3 To enter sample information like sample names and comments, complete
the sample name table.
4 Click the Start button in the upper right of the window to start the chip run.
The incoming raw signals are displayed in the Instrument context.
Contamination of electrodes
C AU T I O N
Leaving the chip for a period longer than 1 hour (e.g. over night) in the Bioanalyzer
may cause contamination of the electrodes.
Immediately remove the chip after a run.
5 After the chip run is finished, remove the chip from the receptacle of the
bioanalyzer and dispose of it according to good laboratory practices.
Agilent DNA 7500 and DNA 12000 Assay Protocol
Starting the Chip Run
5
19