Agilent Technologies DNA 7500 Manual Kit page 22

Table of Contents

Advertisement

6
Checking Your Agilent DNA 7500 and DNA 12000 Assay Results
DNA 7500 and DNA 12000 Ladder Well Results
Figure 3
Major features of a successful ladder run are:
• 12 peaks for the DNA 7500 ladder and 13 peaks for the DNA 12000 ladder
• All peaks are well resolved
• Flat baseline
• Correct identification of both markers
If the electropherogram of the ladder well window does not resemble the
one shown above, refer to the 2100 Expert Maintenance and
Troubleshooting Guide for assistance.
22
DNA 12000 ladder

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dna 12000

Table of Contents