Agilent Technologies DNA 7500 Manual Kit page 5

Table of Contents

Advertisement

Table 1
Physical Specifications
Type
Analysis time
Samples per chip
Sample volume
Kit stability
Kit size
Table 2
Analytical Specifications
Type
Agilent DNA 7500
Sizing range
100 – 7500 bp
Sizing resolution
100 – 1000 bp: 5 %
1000 – 7500 bp: 15 %
Sizing accuracy
± 10 %
Sizing reproducability
5 % CV
Quantitation
20 %
1
accuracy
Quantitation
100 – 1000 bp: 10 % CV
1
1000 – 7500 bp: 5 % CV
reproducibility
1
0.5 – 50 ng/µl
Quantitative range
Maximum salt
250 mM for KCl
15 mM for MgCl
250 mM NaCl
1
Determined using the respective DNA ladder as sample
Agilent DNA 7500 and DNA 12000 Kit
Specification
30 minutes
12
1 µL
4 months (see box for storage temperatures)
12 samples/chip = 300 samples/kit
Agilent DNA 12000
100 – 12000 bp
100 – 1000 bp: 5 %
100 – 12000 bp: 15 %
± 15 %
5 % CV
25 %
100 – 1000 bp: 15 % CV
1000 – 12000 bp: 10 % CV
0.5 – 50 ng/µl
250 mM for KCl
15 mM for MgCl
2
250 mM NaCl
1
2
5

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dna 12000

Table of Contents