ABB FEX300 Operating Instruction page 92

Electromagnetic flowmeter processmaster, hygienicmaster
Hide thumbs Also See for FEX300:
Table of Contents

Advertisement

Parameterization
Menu / Parameter
Easy Setup (continued)
Cable length
Easy Setup / System Zero
Manual
Automatic
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
92
Value range
0.01 ... 200 m
FEX300, FEX500
Description
"Easy Setup" Menu
Enter the signal cable length between the transmitter
and the flowmeter sensor.
For devices with a compact design
(FEP311, FEH311, FEP315, FEH315) 0.01 m must be
entered.
Important (Notice)
The entry is required for FEP500, FEH500 if the
diagnostic functions are to be used.
When using the ScanMaster verification software
you also have to enter the signal cable length.
Submenu "System Zero"
Starts the manual zero adjustment.
Starts the automatic zero adjustment.
Important (Notice)
Prior to starting the zero adjustment, make sure
that:
There is no flow through the flowmeter sensor
(close all valves, shut-off devices, etc.)
The flowmeter sensor is completely filled with
the fluid to be measured.
OI/FEX300/FEX500-EN