ABB FEX300 Operating Instruction page 105

Electromagnetic flowmeter processmaster, hygienicmaster
Hide thumbs Also See for FEX300:
Table of Contents

Advertisement

Menu / Parameter
Input/Output / Logic Setup
DO1 Signal
DO1 Alarm Config
DO1 Action
DO2 Signal
DO2 Alarm Config
DO2 Action
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
OI/FEX300/FEX500-EN
Value range
No Function, F/R Signal,
Alarm Signal, Dual Range 1,
Batch end contact 1
Normally Open, Normally
Closed
No Function, F/R Signal,
Alarm Signal, Dual Range 1,
Batch end contact 1
Normally Open, Normally
Closed
FEX300, FEX500
Description
The menu is displayed only if in the " DO1 / DO2
Function " menu the Logic / Logic function is set. This
menu is not displayed in the factory default setup.
F/R Signal:
The digital output signals the flow direction.
Alarm-Signal:
The digital output functions as alarm output. The
alarm type is set in the DO1 Alarm Config menu.
Dual Range:
The digital output is activated when flow range 2
(Qmax 2) is selected.
Batch end contact:
The digital output is activated when the configured
fill quantity is reached.
Default setting: F/R Signal.
Select the "DO1 Alarm Config" submenu.
The menu is displayed only if in the " DO1 Signal "
menu the Alarm Signal function is set.
Select the switching behavior for the digital output.
Default setting: Normally Open.
Refer to the description "DO1 Signal".
Select the "DO2 Alarm Config" submenu.
The menu is displayed only if in the "DO2 Signal"
menu the "Alarm Signal" function is set.
Refer to the description "DO1 Action".
Parameterization
105