ABB FEX300 Operating Instruction page 117

Electromagnetic flowmeter processmaster, hygienicmaster
Hide thumbs Also See for FEX300:
Table of Contents

Advertisement

Menu / Parameter
Diagnostics / Diagnosis Control / Sensor Measurements
One shot Sensor Mea
Coil Current
Coil Resistor
Coil Voltage
Coil R. Max Alarm
Coil R. Min Alarm
Signal-Cable length
1)
Sensor Temp Value
1)
Adj. Sensor T. Value
1)
Sensor T. Max Alarm
1)
Sensor T. Min Alarm
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
OI/FEX300/FEX500-EN
Value range
Display only
Display only
Display only
0 ... 1000 Ω
0 ... 1000 Ω
0.01 ... 200 m
Display only
-50 ... +200 °C
-50 ... +200 °C
-50 ... +200 °C
FEX300, FEX500
Description
Start the measurement. The measured values for the
start moment are acquired.
Display the coil current.
Display the coil resistance.
Display the coil voltage.
Set the maximum limit value for the coil resistance. In
the case of overshoot an alarm is tripped.
Default setting: 1000 Ω
Set the minimum limit value for the coil resistance. In
the case of undershoot an alarm is tripped.
Default setting: 0 Ω
Enter the signal cable length between the transmitter
and the flowmeter sensor. For devices with a compact
design 0.01 m must be entered.
Default setting: 0 m
Display the sensor temperature.
The sensor temperature must be adjusted according
to the conditions on site. The temperature measured
with a separate thermometer can be entered here.
Set the maximum limit value for the sensor
temperature. In the case of overshoot an alarm is
tripped.
Default setting: -50 °C
Set the minimum limit value for the sensor
temperature. In the case of undershoot an alarm is
tripped.
Default setting: +200 °C
Parameterization
117