ABB FEX300 Operating Instruction page 119

Electromagnetic flowmeter processmaster, hygienicmaster
Hide thumbs Also See for FEX300:
Table of Contents

Advertisement

Menu / Parameter
Diagnostics / Diagnosis Control / Coating Detector (continued)
Coating QE Min Alarm
Coating QE Max Alarm
Diagnostics / Diagnosis Control / Cond. Detection
Cond. Detection
One shot Cond.
Cond. Value
Adj. Cond. Value
Cond. Min Alarm
Cond. Max Alarm
Elec. Imp. E1-GND
Elec. Imp. E2-GND
Elec. Imp Min Alarm
Elec. Imp Max Alarm
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
OI/FEX300/FEX500-EN
Value range
0 ... 100.000
0 ... 100.000
On / Off
5 ... 20,000 µS/cm
5 ... 20,000 µS/cm
5 ... 20,000 µS/cm
Display only
Display only
0 ... 20,000 Ω
0 ... 20,000 Ω
FEX300, FEX500
Description
1)
Set the minimum limit value for the electrode deposit
detection. When the smaller value, QE1 or QE2, is
undershot, an alarm is tripped.
Default setting: 0
Set the maximum limit value for the electrode deposit
detection. When the greater value, QE1 or QE2, is
overshot, an alarm is tripped.
Default setting: 100.000
1)
Activate the "Cond. Detection" function.
Default setting: Off
Important (Notice)
The conductivity measurement function can be
used in the nominal diameter range DN10 ... 300.
For further information read and observe chapter
"Extended diagnostic functions".
The conductivity is measured cyclically at defined
intervals. Here, a current measurement can be
started.
Display the conductivity.
The conductivity must be adjusted according to the
fluid on site. The conductivity measured with a
separate meter can be entered here.
Set the minimum limit value for the conductivity. In the
case of undershoot an alarm is tripped.
Default setting: 5 µS/cm
Set the maximum limit value for the conductivity. In
the case of overshoot an alarm is tripped.
Default setting: 20,000 µS/cm
Current impedance between electrode E1 and GND
(ground potential).
Current impedance between electrode E2 and GND
(ground potential).
Set the minimum limit value for the impedance. In the
case of undershoot an alarm is tripped.
Default setting: 0 Ω
Set the maximum limit value for the impedance. In the
case of overshoot an alarm is tripped.
Default setting: 20,000 Ω
Parameterization
119