ABB FEX300 Operating Instruction page 134

Electromagnetic flowmeter processmaster, hygienicmaster
Hide thumbs Also See for FEX300:
Table of Contents

Advertisement

Extended diagnostic functions
Power Spectrum
Back
Amplitude 1 (n)
Back
The measured values indicate possible disturbances of the device's grounding line at the time of
the test.
No or minor disturbance:
When the power spectrum is above 1000.
When the four measured amplitude values are above 10.
Check the device grounding (!):
When the power spectrum is below 1000.
When the four measured amplitude values are below 10.
134
Grounding Check
335
OK
Grounding Check
2,7
OK
FEX300, FEX500
5. Use
or
to select "Power Spectrum".
6. Use
to call up the parameter for display.
7. Use
or
to select "Amplitude 1 (n)".
8. Use
to call up the parameter for display.
OI/FEX300/FEX500-EN