Menu / Parameter
1)
Diagnostics / Trend
Conductivity
Electrode Coat. QE1
Electrode Coat. QE2
Logtime Interval
Trend-Logging
Diagnostics / Flowrate Alarm
Max. Flowrate Alarm
Min. Flowrate Alarm
Diagnostics / Output Readings
Current
DO1 Frequency
DO2 State
Digital In State
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
OI/FEX300/FEX500-EN
Value range
1 ... 45,000 min.
ON / OFF
0 ... 130 %
0 ... 130 %
mA
Hz
open / closed
open / closed
FEX300, FEX500
Description
The measured values are shown as a line diagram.
When the "Trend-Logging" function is enabled, the
measured values are stored at the defined interval
(cycle time). The last 12 measured values are stored
and indicated as a line diagram. The oldest data
record is overwritten.
Interval for measured value logging.
Activate the "Trend-Logging" function.
When the "Trend-Logging" function is enabled, the
measured values are stored at the defined interval
(cycle time).
The data records can be read out by the
"ScanMaster" verification software and analyzed as
trends.
Set the maximum limit value for the flow.
Set the minimum limit value for the flow.
Display the current values and statuses of the listed
inputs and outputs.
Parameterization
123