Parameterization
Menu / Parameter
Input/Output / Current Output (continued)
Iout at TFE Alarm
Iout Mode
italics = Parameter can only be viewed at the "Advanced" password level.
1) Parameter / menu only available for FEP500 / FEH500.
108
Value range
Off, Q = 0 %, High Alarm, Low
Alarm
4 ... 20 mA, 4 - 12 - 20 mA
FEX300, FEX500
Description
Select the status that the current output shall assume
in the case of a partial filling alarm.
•
Off:
Error is not output at the current output.
•
Q = 0 %:
The current output assumes the value for "No flow
"(4 mA).
•
High Alarm:
The current output assumes the value for "High
Alarm".
•
Low Alarm
The current output assumes the value for "Low
Alarm".
Default setting: „Off".
Select the operating mode for the current output.
•
4 ... 20 mA
-
4 mA = No flow
-
20 mA = Maximum flow
•
4 - 12 - 20 mA
-
4 mA = Maximum reverse flow
-
12 mA = No flow
-
20 mA = Maximum forward flow
OI/FEX300/FEX500-EN