*** What is Time Constant? ***
Time Constant determines the time span of the moving average. When you set the Time
Constant to a larger (longer) value, the readings will become stable. When you select a smaller
(shorter) value, the readings will become more responsive and sensitive to changes in air
velocity. This function is not available in Calculation or Flow Rate Mode.
MODE
TC1
TC5
TC10
20
Normal Measurement
actoolsupply.com

Setting the Time Constant

'12/04/12a15:40:23 ~
0.87f
TCC1
25.3e
64.1g
Time Constant
You can select the Time Constant from 1, 5 or 10seconds.
TC1: Displays the instantaneous value at every 1 second
TC5: Displays moving average for 5 seconds
TC10: Displays moving average for 10 seconds
*The Time Constant is only effective in Normal Mode. When you first turn
on the instrument, it is effective only for the AIR VELOCITY and the AIR
FLOW measurements in Normal Mode. If you need to apply the time
constant to the AIR TEMPERATURE, HUMIDITY and PRESSURE
measurements in Normal Mode, refer to the next section "Changing
Time Constant Application".
How To Take In Measured Data
Data for 1 sec
NORMAL MODE
actoolsupply.com
In NORMAL Mode, you can change the
Time Constant (TC) by pressing
(Measuring Time)
sec.
(Measuring Time)
sec.
Average of 5 sec
(Measuring Time)
sec.
Average of 10 sec
EXPLANATION
Displays the
instantaneous value at
every 1 sec.
Updates every second.
The value displayed is
the average of the past 5
seconds.
Updates every second.
The value displayed is
the average of the past
10 seconds.
.