Kanomax Climomaster S-6531-0E User Manual

Kanomax Climomaster S-6531-0E User Manual

6501 series multifunction anemometer

Advertisement

actoolsupply.com

Advertisement

Table of Contents
loading

Summary of Contents for Kanomax Climomaster S-6531-0E

  • Page 1 actoolsupply.com...
  • Page 2 Kanomax Climomaster 6501-0E Hot-wire Anemometer Kanomax Climomaster 6501-AE Hot-wire Anemometer Kanomax Climomaster 6501-BE Hot-wire Anemometer Kanomax Climomaster 6501-CE Hot-wire Anemometer Kanomax Climomaster S-6531-0E Hot-wire Anemometer Kanomax Climomaster S-6531-1E Hot-wire Anemometer Kanomax Climomaster S-6533-0E Hot-wire Anemometer Kanomax Climomaster S-6533-1E Hot-wire Anemometer...
  • Page 3 actoolsupply.com List of Components ■ Standard Items MODEL Functions - Main Body 6501-00 Main Body with analog 6501-A0 With analog output function output One of Main Body with pressure With pressure measurement 6501-B0 these sensor function Main body with pressure With pressure measurement and sensor and analog 6501-C0...
  • Page 4: Important Safety Information

    actoolsupply.com Important Safety Information The symbols for the warnings used in this manual are defined below: Classifications Danger: To Prevent Serious Injury or Death Warnings in this classification indicate a danger that may result in serious injury or death if not observed. Caution: To Prevent Damage to the Product Warnings in this classification indicate a risk of damage to the product that may void the product warranty if not observed.
  • Page 5 actoolsupply.com Danger Never bring the probe close to a flammable gas atmosphere. The heated sensor may cause a fire or explosion. Forbidden Never touch the sensor. The sensor is heated during operation. Touching the heated sensor may cause burns, and may also damage the sensor itself. Don’t touch Hot surface Do not disassemble or heat the batteries, or put them into a...
  • Page 6 actoolsupply.com Do not use or leave the instrument in a high temperature, high humidity or dusty environment. Do not leave this instrument under direct sunlight for a prolonged period. Forbidden The instrument may not function properly out of the specified operating conditions.
  • Page 7 actoolsupply.com Do not move the main unit and the probe from a cold place to a warm place quickly. It will cause condensation. Even when used in an environment within the specified operating Forbidden temperature and humidity, a sudden temperature change may cause condensation.
  • Page 8: Table Of Contents

    What are Discomfort Index (DI) and Dew Point Temperature (DT)? ......55 About Compensation ..................... 56 § 14 Probe Directivity (Air Velocity) ............... 57 § 15 Trouble Shooting .................. 59 § 16 Warranty and After Service ..............61 Kanomax Limited Warranty ..................61 § 17 Contact Information ................62 actoolsupply.com...
  • Page 9: Part Names And Functions

    actoolsupply.com § 1 Part Names and Functions Main Body Pressure Port ■ Pressure port is only available on MODEL6501-B0/-C0. Probe Socket ■ Analog output is only available on MODEL6501-A0/-C0. Analog Output Terminal LCD Display Serial Communication Terminal DC Input Terminal Power Switch Touch Panel Battery...
  • Page 10: Touch Panel

    actoolsupply.com Touch Panel MENU Key Enters the Main Menu to select a function. *Press the MENU Key to CANCEL out of the current measurement or settings and return to the Main Menu. …………………… Normal Measurement Mode MENUaa1.NORMAL …………… To input duct shape and size aaaaaa2.DUCTaTYPE ………...
  • Page 11: Probe

    actoolsupply.com Probe There are 8 different types of probes available for CLIMOMASTER. The model number and specifications depends on the type of the probe you have. The compatibility feature of the probes allows you to swap out probes freely without having to recalibrate the main unit.
  • Page 12: Probe Cable

    actoolsupply.com Probe Cable The probe can be extended by using a probe cable between the main unit and the probe. Main Body Connector Probe Connector Approx.43 Approx.43 Approx.2000 *Unit: mm Extension Rod (Optional) Straight Cable length: Approx.2000 including the rod Approx.130 Approx.50 Approx.230 to 870...
  • Page 13: Getting Started

    actoolsupply.com § 2 Getting Started Installing Batteries Backside of the body 1. Press down on the battery cover. 2. Slide the cover until it stops. 3. Lift the cover away from the body. 4. Insert the batteries according to the indicated polarity chart.
  • Page 14: Connecting The Probe And The Probe Cable

    actoolsupply.com Connecting the Probe and the Probe Cable This section describes how to connect the probe or the probe cable to the main unit. Make sure that the power is turned OFF when connecting or disconnecting the probe or the probe cable. 1.
  • Page 15: Turning On/Off The Power

    The power switch to turn the power ON/OFF is located at the side of the instrument. When you turn the power on after connecting the probe/probe cable, a test screen with the Kanomax logo, model, and the version of the firmware will be displayed for a few seconds before it switches to the Normal Mode screen.
  • Page 16: Precautions For Measurements

    actoolsupply.com Precautions for Measurements Air Velocity Direction Mark ●MODEL6531/6541/6561: The probe has its own directivity characteristics. Make sure that the direction mark is facing into the airflow. If you are not sure of the airflow direction, slowly rotate Airflow the probe and select the point where you get the maximum velocity reading.
  • Page 17 actoolsupply.com Pressure ●Do not expose the pressure sensor to more than 75kPa of pressure. ■ Pressure sensor is Excess pressure may cause serious damage to the sensor. only available on MODEL6501-B0/-C0. ●The operating temperature must be between 5 C and 40 C or 41 and 104 F when measuring pressure.
  • Page 18: Duct Shape / Size Input

    actoolsupply.com Duct Shape / Size Input Before measuring the volumetric flow rate, the duct shape and size settings must be entered. Up to 25 different duct types can be registered in this instrument. Select a duct type from the registered duct types to measure the air flow rate.
  • Page 19 actoolsupply.com <To Set Duct Size> Registering a Rectangular Duct aaaaaaaDUCTaTYPE to select “3. WIDTH” and 1.ENTRYaNo.aaaaaaaa1 2.SHAPEaaaaRECTANGLE 3.WIDTHoooooooooo800 press to enter the 4.HEIGHTooooooooo500 5.UNIT(mm/inch)aaamm width and press 6.SAVEaINFO to select “4. HEIGHT” and press . Use to enter the height and press The maximum is 9999mm or 999.9 inch.
  • Page 20: Normal Measurement Normal Mode

    actoolsupply.com § 3 Normal Measurement NORMAL MODE This is the mode that you will be in, when you first turn on the instrument. In this mode you cannot save any data. The display is updated every 1 second. To move to NORMAL Mode from MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION...
  • Page 21 actoolsupply.com Hold the Reading While measuring (in NORMAL Mode), '12/04/12a15:40:23 ~ ■ Humidity is only press the key to hold the current available on 0.87f TCC1 MODEL6531/6533. reading. This is applicable in both the 25.3e ■ Air temperature is flow rate and pressure measurement 64.1g only available on screens as well.
  • Page 22: Setting The Time Constant

    actoolsupply.com Setting the Time Constant In NORMAL Mode, you can change the '12/04/12a15:40:23 ~ ■ Humidity is only available on 0.87f TCC1 Time Constant (TC) by pressing MODEL6531/6533. 25.3e ■ Air temperature is 64.1g only available on MODEL6531/6541 Time Constant /6542/6533/6543/ You can select the Time Constant from 1, 5 or 10seconds.
  • Page 23 actoolsupply.com Changing the Time Constant Application ■ Pressure is only When you first turn on the instrument, the Time Constant is only available on effective for AIR VELOCITY and AIR FLOW in Normal Mode. If you MODEL6501-B0 want to make it effective for AIR TEMPERATURE, HUMIDITY and /-C0.
  • Page 24: Measuring Max, Min, And Mean: Calculation Mode

    actoolsupply.com § 4 Measuring Max, Min, and Mean: CALCULATION Mode CALCULATION Mode will automatically calculate the maximum, minimum and mean of the measured data. The 1 trial The 2 trial The N trial TRIAL(1) TRIAL(2) TRIAL(N) …………… Sampling Time Measuring Time……Sampling Time x N (Number of Trial) Press to enter the Main Menu.
  • Page 25 actoolsupply.com <Setting the CALCULATION MODE> CALCULATE R20000/20000 Select “1. MODE“ and press 1.MODE11111111AVERAGE to select AVERAGE or 2.SAMPLING(TIME)0012s 3.SAMPLING(n)110050 INSTANT and press 4.DATA2STORAGE2?22YES 5.SET2TO2START <Setting the SAMPLING TIME> to select CALCULATE R20000/20000 “2. SAMPLING (TIME)” and press 1.MODE11111111AVERAGE 2.SAMPLING(TIME)0012s to select SAMPLING TIME 3.SAMPLING(n)110050 4.DATA2STORAGE2?22YES (1 to 999 sec) and press...
  • Page 26 actoolsupply.com <READY> The instrument is ready for measuring. ■ Humidity is only '12/04/12a15:40:23 ~ available on 0.87f MODEL6531/6533. RDY. Press to change the applied 25.3e N2221 ■ Air temperature is parameters. (Air Velocity, Air Temp, /2250 only available on 64.1g Humidity –...
  • Page 27: Measurement Of Air Flow Flow Rate Mode

    actoolsupply.com § 5 Measurement of Air Flow FLOW RATE Mode CLIMOMASTER features an accurate Flow Rate Mode which corresponds to the industry measurement standards such as ASHRAE. SINGLE FLOW RATE Mode This mode is useful for performing a measurement of the airflow of duct (i.e.
  • Page 28 actoolsupply.com Air Flow: Air Volume per Time Unit [m /min, m /h, ft /min, ft Air Flow (Q) = Average Air Velocity (U) x Cross Sectional Area (A) Measurement Point r1 r2 r1=0.316R r2=0.548R r3=0.707R r4=0.837R r5=0.949R Rectangular Duct Circular Duct MENUaa1.NORMAL Press to enter the Main Menu.
  • Page 29: Data Storage

    actoolsupply.com <Setting the No. TRIAL> to select SINGLE (R20000/20000 1.SAMPLING(TIME)001os “2. SAMPLING(n)”and press 2.SAMPLING(n)110050 3.MEAS.2POINT220010 to select No. TRIAL (1 to 4.DATA2STORAGE2?22YES 5.DUCT2ENTRY2NO.2221 6.SET2TO2START 9999) and press <Setting MEAS. POINT> SINGLE (R20000/20000 to select 1.SAMPLING(TIME)001os “3. MEAS. POINT” and press 2.SAMPLING(n)110050 3.MEAS.2POINT220010 to select the number of...
  • Page 30 actoolsupply.com <Ready> The ready screen will be displayed. '12/04/12a16:11:59 ~ while pressing --6037a RDY. P 2221 change the entry no. of the duct type. 25.1e N2221 D2223 63.7g C2300 Press to start measuring. Display Icons 1. RDY: Current Status (READY/SAMPLE) 2.
  • Page 31: Multi Flow Rate Mode

    actoolsupply.com MULTI FLOW RATE Mode MULTI FLOW RATE is very similar to the SINGLE FLOW RATE MODE, except it adds an additional parameter for the number of locations. This allows you to take several air flow measurements (i.e. traverses) and get a total average, for example measuring flow rate from multiple ducts to get an overall average.
  • Page 32 actoolsupply.com <Save the Setting> to select MULTI (R20000/20000 “6. SET TO START” and press 1.SAMPLING(TIME)001ps 2.SAMPLING(n)110050 3.MEAS./LOC.oo010/010 4.DATA2STORAGE2?22YES 5.DUCT2ENTRY2NO.2221 6.SET2TO2START If you press before saving the settings you will return back to the Main Menu and any setting changes you made will not be saved. <Ready>...
  • Page 33: Store And Recall Measured Data

    actoolsupply.com § 6 Store and Recall Measured Data Data which can be stored in the memory of the instrument is shown below. What can be Stored Measuring MODEL CALCULATION Mode FLOW RATE Mode Mode Display 6531 6533 Stored V, W V, W V, W Parameters...
  • Page 34 actoolsupply.com <Setting the PAGE> to select the page that CALCULATION you would like to recall and press 22222DAPAGEOSET ……………………… Page Number PAGE:0001 ………… Measured Mode (CALCULATION (A): AVERAGE/(I): INSTANT ) MODE:CALCULATION(A) ………………… Measured Date (Year/Month/Day) DATE:2012/04/12 ………………… Measured Time (Hour : Minute : Second) TIME:2213:40:53 ………...
  • Page 35 actoolsupply.com <Setting the CALCULATION RANGE> You can select the range of calculation. If the data range is correct, press to calculate the range of data displayed on the screen. If the range is not correct, press START:00100END:050 NUM.0m/s0000°C000%RH display a cursor on “START”. 001000.810025.40064.7 002000.950025.40064.7 003000.980025.60064.9...
  • Page 36: Print Out

    actoolsupply.com § 7 Print Out Preparation for Printing Serial Communication Connect the printer cable to the Serial Terminal Communication terminal located on the side of the instrument to print out measurement data. <Requirements> ●Printer (optional)……Recommended: Seiko Instruments Model DPU-S245 ●Printer Cable (optional) <Check the BAUD RATE>...
  • Page 37: Print Out In Calculation Mode And Flow Rate Mode

    actoolsupply.com Print Out in CALCULATION Mode and FLOW RATE Mode <Measurement and Calculation Mode (CALCULATION)> Press after the measurement and aNEXCALCULATION calculation has been performed to print 102.76f out the result. 081.43f 060.81f <Airflow Measurement Mode (FLOW RATE)> Press after the measurement and aNEXT1FLOW2RATE calculation has been performed to print 106433a...
  • Page 38: Print Out Stored Data(Single Page

    actoolsupply.com Print Out Stored Data(Single Page) Press to enter the Main Menu. MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION to select “5. DATA OUTPUT” aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT and press aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO DATA OUTPUT to select 1.DISPLAY 2.PRINTER(SINGLE) “2. PRINTER(SINGLE)” and press 3.PRINTER(MULTI) to select the page that you would to print out and press 22222DAPAGEOSET ………………………...
  • Page 39 actoolsupply.com AaaaaPRINT OUTPUT Press to select the contents of the 1.RESULT Print Out. 2.DATA 3.ALL to select from items 1 to 3 and press to print out. Please refer to the printing example shown below. 1. RESULT… Conditions and Calculation Results 2.
  • Page 40: Print Out Stored Data(Plural Pages

    actoolsupply.com Print Out Stored Data(Plural Pages) Press to enter the Main Menu. MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION to select “5. DATA OUTPUT” aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT and press aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO to select DATA OUTPUT 1.DISPLAY “3. PRINTER(MULTI)” and press 2.PRINTER(SINGLE) 3.PRINTER(MULTI) ooDATAoOUTPUT(MULTI) to select 1.SELECToSTART:o0001 oooooooooENDoo:o0003 oooooooooSEToo:oooNO...
  • Page 41: Digital Output

    actoolsupply.com § 8 Digital Output Preparation Serial Communication Terminal You can download the data stored in the CLIMOMASTER to your PC, by connecting the CLIMOMASTER to your PC with the USB cable. <Requirements> ● Personal Computer ● USB Cable (supplied) ●...
  • Page 42 actoolsupply.com Transmission of On-Time Data Measured at every 1sec <Enter the Number of Data Needed> 1. Type “D**** ”. *Type in the number of data(4 digits) after “D”. 2. After “AD”, the data will be displayed. 3. Each data represents 1 second of measurement. If you ask for 20 data, it takes approximately 20 seconds to display.
  • Page 43 actoolsupply.com Transmission of Stored Memory Example <To Download the Duct Setting> 1. Type “K ” 2. After “AK”, the data will be displayed. 3. All stored duct settings (1 through 25) will be downloaded. Output Content Entry No. (Data Location); Width; Height; Diameter; Units of Duct Size Example <To Download the Page Number>...
  • Page 44 actoolsupply.com <To Download the Measuring Condition of Stored Data> 1. Press “M**** ”. *Type in the desired page number(4 digits) after “M”. 2. After “AM”, the data will be displayed. Output Content Measuring Parameter WTH: Flow Rate; Temperature; Humidity Flow Rate; Temperature VTH: Velocity;...
  • Page 45: Analog Output

    actoolsupply.com § 9 Analog Output Analog Output 1. Data Update……every 0.1 sec ■ Analog output is only Terminal (Except Humidity; updated every 1sec) available on 2. Load Impedance……Above 5KΩ MODEL6501-A0/-C0. 3. Output Voltage……DC 0 to 1V For analog output, you must select one setting from the table below.
  • Page 46 actoolsupply.com Press to enter the Main Menu. MENUaa1.NORMAL aaaaaa2.DUCTaTYPE to select “7. UTILITY”. aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT Press aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO to select aaaaaaaUTILITY oooooooUTILITY “3. ANALOG OUTPUT”. 1.TIME AJUST 1.TIMEoADJUST 2.UNIT SELECT 2.UNIToSELECT 3.ANALOG OUTPUT 3.ANALOGoOUTPUT Press 4.TC(T.H.P) SET 4.TC(T.H.P)oSET 5.BACKLIGHToON/OFF <To Select Data>...
  • Page 47: Other Settings

    actoolsupply.com § 10 Other Settings Date MENUaa1.NORMAL Press aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION to select “7. UTILITY” and aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR press aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO oooooooUTILITY to select “1. TIME ADJUST” 1.TIMEoADJUST 2.UNIToSELECT and press 3.ANALOGoOUTPUT 4.TC(T.H.P)oSET 5.BACKLIGHToON/OFF AaaaaTIME2ADJUST *The style of date to select "1. STYLE" or 1.STYLE:222JP displayed on the 2.DATE2:22012/04/12...
  • Page 48: Units And Baud Rate

    actoolsupply.com Units and Baud Rate <Units Conversion Table> Velocity 1m/s = 196 FPM Temperature T( F) = 1.8×T(°C)+32 Flow Rate /h = 35.32ft Press to enter the Main Menu. MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION to select “7. UTILITY”. aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT Press aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO oooooooUTILITY...
  • Page 49: Settings For Lcd Backlight

    actoolsupply.com Settings for LCD Backlight MENUaa1.NORMAL Press aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION to select “7. UTILITY”. aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR Press aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO oooooooUTILITY to select 1.TIMEoADJUST “5. BACKLIGHT ON/OFF”. 2.UNIToSELECT 3.ANALOGoOUTPUT Press 4.TC(T.H.P)oSET 5.BACKLIGHToON/OFF to select “1. BACKLIGHT”. ooooLCDoBACKLIGHT Press 1.BACKLIGHToON to select ooooFORoooooooo01omin BACKLIGHT ON/OFF.
  • Page 50: To Delete Data

    actoolsupply.com To Delete Data Press to enter the Main Menu. to select “6. DATA CLEAR” MENUaa1.NORMAL aaaaaa2.DUCTaTYPE and press aaaaaa3.CALCULATION aaaaaa4.FLOWaRATE to select “1. CLEAR” (delete aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY the selected data only) and press aaaaaa8.PRESSUREaZERO AaaaaDATARCLEAR ……Starting Page Number 1.CLEARLSTART:000001 ……...
  • Page 51: To Delete All The Data

    actoolsupply.com To Delete All the Data Press to enter the Main Menu. Use t to select “6. DATA CLEAR”. MENUaa1.NORMAL aaaaaa2.DUCTaTYPE aaaaaa3.CALCULATION Press aaaaaa4.FLOWaRATE aaaaaa5.DATAaOUTPUT aaaaaa6.DATAaCLEAR aaaaaa7.UTILITY aaaaaa8.PRESSUREaZERO to select “2. ALL CLEAR”. Press AaaaaDATARCLEAR 1.CLEARLSTART:000001 00000000END00:000001 00000000SET00:0020NO ……All Delete (YES or NO) 2.ALL2CLEAR0000000NO ……Remaining Memory / Total Memory 000000000R19950/20000...
  • Page 52: Contrast Adjustment

    actoolsupply.com Contrast Adjustment If the LCD display of the CLIMOMASTER is dark light, there adjustment dial at the back, bottom of the unit, inside the battery cover. Contrast Adjusting Dial You can adjust the contrast by using a precision screwdriver (-) (0.9 to 1.5mm).
  • Page 53: How To Clean A Probe

    actoolsupply.com § 11 How to Clean a Probe Dust or particles attached to the velocity sensor would alter the amount of heat diffusion, which leads to a less precise reading. Also, deformation or clogging of the protective mesh around the sensor of the probe would also affect the accuracy of the instrument.
  • Page 54: Specification

    actoolsupply.com § 12 Specification CLIMOMASTER Air Velocity Meter Product 6501 Model No. of the Main Body 6541 6561 6542 6551 6552 6543 Model No. of the Probe 6531-21 6533-21 Measuring Object Clean air flow Measuring Range 0.01 to 30.0, 0.01 to 50.0 (6561-21 only) 0.01 to 5.00 0 to 9.99: 0.01 Resolution...
  • Page 55: Principle Of Measurement

    actoolsupply.com § 13 Principle of Measurement Principle of Hot-wire Anemometer The principle of the thermal Probe is based on a heated element from which Current heat is extracted by the colder impact flow. The temperature is kept Cooling constant via a regulating switch. The Wind controlling current is directly -off...
  • Page 56 actoolsupply.com Measurement of Wind-Temperature (MODEL 6531/6541/6542/6533/6543/6561) Wind The coefficient of resistance of the temperature sensor has a Wind direct proportional relationship with the temperature. We can measure the wind temperature by adjusting the sensor’s temperature to the wind-temperature, and measuring its Temp Sensor (Platinum Resistor) coefficient of resistance.
  • Page 57: What Are Discomfort Index (Di) And Dew Point Temperature (Dt)

    actoolsupply.com What are Discomfort Index (DI) and Dew Point Temperature (DT)? **What is DI and DT?** ● Discomfort Index (DI) Discomfort Index is an index to show how uncomfortable it is in the summer time. CLIMOMASTER use the formula, shown below, which is used by Meteorological Agencies.
  • Page 58: About Compensation

    Please contact your local Kanomax representative for details. Principle of Measurement actoolsupply.com...
  • Page 59: Probe Directivity (Air Velocity)

    actoolsupply.com § 14 Probe Directivity (Air Velocity) MODEL 6531-21 / 6541-21 / 6561-21 Horizontal Wind Direction Indicator 90° -180° 0° -90° At 5m/s Vertical Wind Direction Indicator -90° 90° At 5m/s actoolsupply.com...
  • Page 60 actoolsupply.com Probe Directivity (Air Velocity) MODEL 6542-21 Horizontal Vertical 0° -180° 90° -90° 90° 0° -90° -180° At 1m/s Probe Directivity (Air Velocity) M O D E L 6 5 3 3 - 2 1 / 6 5 4 3 - 2 1 / 6 5 5 1 - 2 1 / 6 5 5 2 - 2 1 Vertical Horizontal 0°...
  • Page 61: Trouble Shooting

    The probe or the probe cable may be damaged. Page 62  Contact your local Kanomax representative for repair or replacement. instrument The probe might be not facing into the wind. reading the correct air ...
  • Page 62 actoolsupply.com Output check (1) - Print out Problem Possible cause(s) / Solution(s) Reference The printer is not connected properly. Page 34  Confirm the printer connection. The Baud Rate might not be set properly.  Check both the instrument and printer Page 46 settings.
  • Page 63: Warranty And After Service

    Warranty and After Service Kanomax Limited Warranty The limited warranty set forth below is given by KANOMAX JAPAN, Inc. (hereafter referred to as “KJI”) with respect to the KANOMAX brand anemometer, and its attachment parts including probe and other accessories (hereafter referred to as “PRODUCT”) purchased directly from KJI or from and authorized distributor.
  • Page 64: Contact Information

    Contact Information No copying, distribution, publication, modification, or incorporation of this document, in whole or part, is permitted for commercial purposes without the express written permission of Kanomax. The contents of this document may be changed without prior notice. 01002/12.04 Contact Information actoolsupply.com...