7.7 Pulse output

Problem
No output.
Output signal levels are
different.
The output value is over

7.8 Alarm output

Problem
No output.
Or the output timing is
unknown.
Output
signal
levels
different.
Alarm is always output.

7.9 Date and Time Setting

Problem
Date/time
is
significantly
off.
Or reset continuously
after setting.
Possible cause(s) / Solution(s)
The plug is not properly inserted to the output terminal.
Make sure using a dedicated output cable.
Make sure to use the voltage and current within the rated
range.
Make sure the polarity (+, -) is correct.
The instantaneous value exceeds a concentration
equivalent to 10,000 CPM.
Possible cause(s) / Solution(s)
The plug is not properly inserted to the output terminal.
Make sure using a dedicated output cable.
Make sure the alarm setpoint is set correctly.
Make sure the alarm is set to ON.
The alarm is only output during the start to the end of
the measurement.
Make sure to use the voltage and current within the rated
are
range.
Make sure the polarity (+, -) is correct.
The instantaneous value exceeds a concentration
equivalent to 10,000 CPM.
Possible cause(s) / Solution(s)
The backup battery is going to run out.
We recommend replacing the backup battery at the time
of periodic calibration (1 year). Please contact our
service center or distributor.
The battery life varies depending on the frequency of
use, operating environment. Since the backup battery
is a primary battery, it cannot be charged when the main
unit is powered on.