Analog Output

1. Data Update......every 0.1 sec
Analog output is only
available on
2. Load Impedance......Above 5KΩ
MODEL6501-A0/-C0.
3. Output Voltage......DC 0 to 1V
For analog output, you must select one setting
from the table below.
(There is no analog output for Flow Rate)
Temperature
To change the setting
For the output range, the low end will be set at 0V and the high end
of Time Constant,
will be set at 1V. The voltage is linear.
refer to "To change
The Time Constant will be in effect.
Time Constant" on
P.20.
Ways To Take In Measured Data (Analog Output)
Mode
0
TC1
0
TC5
0
TC10
actoolsupply.com
(Except Humidity; updated every 1sec)
Parameter
Output Range
0 to 5
0 to 10 m/s
0 to 30 m/s
0 to 50 m/s
Velocity
(U)
0 to 1000 FPM
0 to 2000 FPM
0 to 6000 FPM
0 to 10000 FPM
0 to 50
Humidity
(H)
0 to 100 %RH
-20 to 30 °C
0 to 50 °C
0 to 100 °C
(T)
-4 to 86
32 to 122
32 to 212
Pressure
-2 to +2 kPa
(P)
-5 to +5 kPa
Measuring time
0.5
1.0
1.5
2.5
5.0
7.5
5
10
15
actoolsupply.com
Analog Output
Terminal
Conversion Formula
m/s
U= 5×V
U= 10×V m/s
U= 30×V m/s
U= 50x V m/s
U= 1000×V FPM
U= 2000×V FPM
U= 6000× V FPM
U= 10000×V FPM
%RH
H= 50×V
H= 100×V %RH
T= 50×V-20 °C
T= 50×V °C
T= 100×V °C
o
F
T= 90×V - 4
o
F
T= 90×V+32
o
F
T= 180×V+32
P= 4×V- 2
P= 10×V- 5 kPa
2.0sec
Average of 1 sec
10sec
Average of 5sec
20sec
Average of 10sec
(V: voltage)
m/s
%RH
o
F
o
F
o
F
kPa
Explanation
Take 10
measurements each
second and indicate
the average as an
instantaneous value
at every 1 sec.
Output the average
measured value for
5 seconds at every
0.1 second. Data
shifts by 0.1 second.
Output the average
measured value for
10 seconds at every
0.1 second. Data
shifts by 0.1 second.
43