Measurement of Air Flow FLOW RATE Mode
actoolsupply.com
MULTI
(R20000/20000
1.SAMPLING(TIME)001ps
2.SAMPLING(n)110050
3.MEAS./LOC.oo010/010
4.DATA2STORAGE2?22YES
5.DUCT2ENTRY2NO.2221
6.SET2TO2START
before saving the settings you will return back to the
'12/04/12a16:11:59 ~
---502a
RDY.
P2221
27.1e
N2221
L2221
62.9g
C2300
Display Icons
1. RDY: Current Status (READY/SAMPLE)
2. P 1: Current Number of Points
3. N 1: Current Number of Trial
4. L 1: Current Number of Location
5. C300: Duct Shape and Size
(In this case, the duct is circular and 300mm in diameter.)
For the rectangular duct, W*** and H*** will be displayed.
aNEXT1FLOW2RATE
106433a
MAX
085134a
AVG
066797a
MIN
Press
to print the results, when a printer is connected.
Press
to return to the Main Menu.
Press
to return to the setup screen of FLOW RATE Mode.
*When "DATA STORAGE" is set to YES, values of each location will
be stored.
actoolsupply.com
Use
to select
"6. SET TO START" and press
The ready screen will be displayed.
Use
, while pressing
change the entry number of the duct
type. Press
to start measuring.
Follow the same procedure as SINGLE
mode to measure.
After all the measurements at all
locations are completed, the
calculation results will be displayed.
Press
to display the results of
each item in the following sequence:
Flow rate→Air Temp→Humidity→
Air velocity
.