Pressure
●Do not expose the pressure sensor to more than 75kPa of pressure.
Pressure sensor is
Excess pressure may cause serious damage to the sensor.
only available on
MODEL6501-B0/-C0.
●The operating temperature must be between 5
and 104
operate properly outside of this temperature range.
●Make sure to perform a zero adjustment before measuring
pressure. When performing a zero adjustment, leave both pressure
ports (+) and (-) open.
●The zero adjustment range is – 0.5 to 0.5 kPa. If the reading is out
of this range, an error message will be displayed.
■Zero Adjustment Procedure
MENUaa1.NORMAL
aaaaaa2.DUCTaTYPE
aaaaaa3.CALCULATION
aaaaaa4.FLOWaRATE
aaaaaa5.DATAaOUTPUT
aaaaaa6.DATAaCLEAR
aaaaaa7.UTILITY
aaaaaa8.PRESSUREaZERO
aaaaPRESSUREaZERO
1.ZEROaADJUSTaaaYES
2.SAVEaINFO
aaaaPRESSUREaZERO
1.ZEROaADJUSTaaaYES
2.SAVEaINFO
aaaaPRESSUREaZERO
1.ZEROaADJUSTaaaYES
2.SAVEaINFO
*In order to take
an accurate
measurement,
make sure that the
tube is properly
secured without
any leakage, or
bends.
actoolsupply.com
o
F when measuring pressure. The instrument may not
-0.00h
00.00h
00.00h
If you press
before you save, you will go back to the Main Menu
without completing the zero adjustment.
Connecting the Pressure Tube
Connect the pressure tube to the (+) or (-)
port as you see on the left.
Connect/insert the other end of the tube to the
measuring hole of the duct. If the pressure
that you are going to measure is positive,
connect the tube to (+) port. If it is negative,
connect the tube to (-) port.
actoolsupply.com
Press
to enter the Main Menu.
Use
to select "8. PRESSURE ZERO"
and press
.
Use
to select "1. ZERO ADJUST"
and press
.
Use
to select "YES".
Press
to display 0.00 kPa.
Use
to select "2. SAVE INFO".
Press
. This will complete the zero
adjustment and you are back in Main
Menu.
o
o
o
C and 40
C or 41
F
15