Specification - Kanomax Climomaster S-6531-0E User Manual

6501 series multifunction anemometer
Table of Contents

Advertisement

§ 12

Specification

Product
Model No. of the Main Body
Model No. of the Probe
Measuring Object
Measuring Range
Resolution
Air Velocity
[m/s]
Accuracy
Response Time
[sec.]
Measuring Range
Resolution
Air
Accuracy
Temperature
o
[
C]
Response Time
[sec.]
Measuring Range
Resolution
*1
Humidity
Accuracy
[%RH]
Response Time
[sec.]
Measuring Range
Resolution
2
Pressure
Accuracy
[kPa]
Response Time
[sec.]
Temperature Compensation
[m/s]
Functions
Output
Power
Battery Life
Operating Environment
Storage Environment
Weight
Accessories
Options
*1: Air temperature is only available on MODEL6531/6541/6542/6533/6543/6561
*2: Humidity is only available on MODEL6531/6533
*3: Pressure is only available on MODEL6501-B0/-C0 , Analog Output is only available on MODEL6501-A0/-C0
*4: Optional.
52
Specification
actoolsupply.com
CLIMOMASTER Air Velocity Meter
6541
6561
6531-21
-21
-21
0.01 to 30.0, 0.01 to 50.0 (6561-21 only)
0 to 9.99: 0.01
10 to 30:
0.1
30 to 50:
0.1 (6561-21 only)
±2% of reading or ±0.015 (±3FPM),
whichever is greater
Approx. 1
Approx. 4
(at 1m/s, 90% response time)
-20.0 to 70.0
0.1
±0.5
Approx. 30
(at 1m/s, 90% response time)
2.0 to 98.0
0.1
2 to 80:
±2.0
80 to 98:
±3.0
Approx.
15
±3% of reading or ±0.01
±5% of reading or ±0.015 (±3FPM), whichever is greater
(in the temperature range of 5 to 60.0
Readings Hold, Max Hold, Time Constant (1, 5, 10 sec.), Battery Level
Indicator (5 Levels),
Unit selection (Air Velocity: m/s or FPM, Flow Rate: m
3
3
ft
/min, or ft
/h, Temperature
*3
Pressure
: kPa or Pa)
Max., Min., and Average Calculation, Sampling Time: 1 to 999 sec.
No. of trails: 1 to 9999, Store up to 25 Duct Sizes (Square/Round, Size
Range: 1 to 9999mm or 0.1 to 999.9inch(diameter, height, width))
Max.Data Storage: 20000
Digital Output: USB (when connected to a printer, automatically switch
to RS-232C, Baud Rate 4800, 9600, 19200 and 38400 bps)
*3
Analog Output
: DC 0 to 1V (select one from Air Velocity,
*1
Temperature
, Humidity
6 × AA Batteries , AC Adaptor
Approximately 10 continuous hours (at air velocity 5m/s, 20
alkaline batteries)
o
Main Body: 5 to 40
C (41 to 104
o
Probe: -20 to 70
C (-4 to 158
o
o
-10 to 50
C (14 to 122
Approximately 400g (batteries included)
Carrying case, Operation manual, 6×AA Batteries, Probe cable (2m),
USB cable
Spare probe, Analog output, Pressure sensor, Extension rod, Printer,
Printer cable, Software (for Windows), AC adaptor, Hands-free Case
actoolsupply.com
6501
6542
6551
6552
-21
-21
-21
Clean air flow
Approx. 7
-5.00~+5.00
0.01
Approx. 1
o
C)
*1
o
o
:
C or
F, Humidity
*2
*3
and Pressure
)
*4
: AC 100 to 240V (50/60Hz)
o
F) with no visible condensation
o
F) with no visible condensation
F) with no visible condensation
6543
6533-21
-21
0.01 to 5.00
0.01
0.01 to 0.99:±0.02
0.99 to 5.00:
±2% of reading or
±0.015 (±3FPM),
whichever is greater
Approx. 7
-20.0 to 70.0
0.1
±0.5
Approx. 30
(at 1m/s, 90%
response time)
2.0 to 98.0
0.1
2 to 80:
±2.0
80 to 98:
±3.0
Approx.
15
3
3
/min, m
/h,
*2
: %RH, Static
o
C, with

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents