Kanomax Climomaster S-6531-0E User Manual page 62

6501 series multifunction anemometer
Table of Contents

Advertisement

Output check (1) - Print out
Problem
The printer doesn't
operate.
Cannot printout the
display
Want to cancel printing
Output check (2) - Digital output
Problem
PC doesn't load the
measuring data.
Output check (3) - Analog output
Problem
No data is generated.
Data is tiered.
The output data is not
correct.
60
Trouble Shooting
actoolsupply.com
Possible cause(s) / Solution(s)
The printer is not connected properly.
 Confirm the printer connection.
The Baud Rate might not be set properly.
 Check both the instrument and printer
settings.
Make sure to use the optional printer (Model
DPU-S245)
and
6000-31).
Confirm the connection procedure.
 Connect the printer to the printer port,
then turn on the instrument and printer.
Display is not frozen.
 1. Press
to hold the display.
2. Press
to print out.
You can't cancel print while it is operating.
Possible cause(s) / Solution(s)
Confirm
the
USB
different from printer connection.
The Baud Rate might not be set properly.
 Check both the instrument and PC
settings.
Confirm the communication command (ex.
Hyper Terminal command).
Possible cause(s) / Solution(s)
Check the connection and confirm the
polarity.
The HOLD function might be active.
 Press
to cancel the HOLD function.
Data is outputted every one second.
Confirm the setting for analog output.
Confirm the output range.
The load impedance might not be optimum.
Set the load impedance larger than 5k
actoolsupply.com
printer
cable
(Model
connection, which
Ω.
Reference
Page 34
Page 46
Page 34
Page 34
Page 34
Page 34
Reference
is
Page 39
Page 34
Page 39
Reference
Page 43
Page 43
Page 43
Page 43
Page 43
Page 43

Hide quick links:

Advertisement

Table of Contents
loading