Appendix A : Error Codes - Laversab 6300 User Manual

High accuracy automated pressure controller
Hide thumbs Also See for 6300:
Table of Contents

Advertisement

6300 Rev H User's Manual
Error code
1
Pitot Actual value exceeds knots limit or Mach limit
2
Pitot Rate Actual value exceeds knots/min limit
3
Static Actual value exceeds Feet limit
4
Static Rate Actual value exceeds Feet/min limit
5
Pitot control transducer error
6
Static control transducer error
7
Pitot leak-rate exceeds 50 knots/min; automatically returns Pitot to control
8
Static leak-rate exceeds Static Rate target; automatically return Static to control
9
Knots limit will soon be exceeded in Leak mode; returned to control mode
10
Warning: Loss of AC power. Operating on battery
11
Loss of data from Pitot or Static transducer
12
Pitot target exceeds knots limit
13
Pitot target exceeds Mach limit
14
Pitot Rate target exceeds knots/min limit (or 30 inHg/min for pressure units)
15
Static target exceeds Feet limit
16
Static Rate target exceeds Feet/min limit (or 30 inHg/min for pressure units)
17
Leaving Pitot in Measure mode may cause knots limit to be exceeded
18
Leaving Pitot in Measure mode may cause Mach limit to be exceeded
19
Serial interface error: invalid command or invalid value or syntax error
20
Calibration incomplete (Status code returned after CC command)
21
Pitot calibration complete (Status code returned after CC command)
22
Static calibration complete (Status code returned after CC command)
23
Both Pitot & Static calibration complete (Status code returned after CC command)
24
Password not set (either for Limits or Calibration)
Cannot enter Calibration unless Pitot is in QcinHg and Measure mode, and Static is in
25
inHg and Measure mode
125-9106A

APPENDIX A : ERROR CODES

Description
Page 80