6300 Rev H User's Manual
9.
Enter '0.000' (or simply 0) followed by
replaced by '0.000'. Press
'0.000' on this screen, as the Pitot vacuum point. This indicates to the 6300 that the differential
pressure on the Pitot output is currently exactly 0.000 inHg.
10. Generate exactly QcinHg=30.000 and PsinHg=16.000. Wait for a minimum of 5 minutes to allow
temperature effects to stabilize.
11. Press '2' on the Remote to enter the Pitot full-scale point. The screen will appear as follows.
12. Enter '30.000' (or simply 30) followed by
replaced by '30.000'. Press
see '30.000' on this screen, as the Pitot full-scale point. This indicates to the 6300 that the
differential pressure on the Pitot output is currently exactly 30.000 inHg.
13. Generate exactly QcinHg=0.000 and PsinHg=32.000. Wait for a minimum of 5 minutes to allow
temperature effects to stabilize.
14. Press '4' on the Remote to enter the Static full-scale point. The screen will appear as follows.
15. Enter '32.000' (or simply 32) followed by
replaced by '32.000'. Press
see '32.000' on this screen, as the Static full-scale point. This indicates to the 6300 that the
absolute pressure on the Static output is currently exactly 32.000 inHg.
125-9106A
Figure 6.5
ENTER
on the keypad. The highlighted field will be
CLR
to return to the point-selection screen (Figure 6.3). You should see
Figure 6.6
ENTER
CLR
to return to the point-selection screen (Figure 6.3). You should
Figure 6.7
ENTER
CLR
to return to the point-selection screen (Figure 6.3). You should
on the keypad. The highlighted field will be
on the keypad. The highlighted field will be
Page 66