Appendix A: Error Codes - Laversab 6300 User Manual

Hide thumbs Also See for 6300:
Table of Contents

Advertisement

All errors are displayed on line 4 of the display when the 6300 is in the Main screen. An
error must be acknowledged before any other key entries are accepted. The error may be
acknowledged either by pressing the CANCEL or the CLR key.
ERROR CODES
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
APPENDIX A
ERROR CODES
DESCRIPTION
Pressure exceeded 110 %FS on Pt output
Pressure exceeded 110 %FS on Ps output
Pt target value exceeds Qc limits
Pt target value exceeds Mach limit
Ps rate target exceeds Rate limit
Ps target exceeds altitude limits
Pt actual value exceeds Qc limit
Pt actual value exceeds Mach limit
Rate exceeds limits.
Unused
Invalid Profile point
Ps control-transducer error
Pt control-transducer error
Ps actual value exceeds limit
Internal Use only
Internal Use only
If Pt is left in Measure mode, Qc limit will soon be exceeded
If Pt is left in Measure mode, Mach limit will soon be exceeded
Loss of Actual Data
62