6300 Rev H User's Manual
The table shown below provides the pressure values for QcinHg and PsinHg that need to be generated
for each point of the verification process. It also includes, for each point, the offset commands (followed
by specific values) that may be entered through the RS232 port. This table is referenced in the procedure
that follows.
Point
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
Offsets are entered through Hyperterminal as follows:
For Static offsets:
For Pitot offsets:
where n is the offset number between 1 and 9, and 'xxxx' is the positive or negative offset value equal to
(Actual value – Pressure Standard value).
For example, at point 2 (Pressure Standard = 3.200 PsinHg), if the 6300 Static Actual value is 3.198, the
offset value will be (3.198 – 3.200 = -0.002), and the command sent to the 6300 would be:
At point 2, if the 6300 Static Actual value is 3.201, the offset value will be (3.201 – 3.200 = 0.001), and the
command sent to the 6300 would be:
125-9106A
Verification points
QcinHg
Pitot offset
0.000
0.000
0.000
0.000
0.000
0.000
No offset
3.000
P1 (10%)
6.000
P2 (20%)
9.000
P3 (30%)
12.000
P4 (40%)
15.000
P5 (50%)
18.000
P6 (60%)
21.000
P7 (70%)
24.000
P8 (80%)
27.000
P9 (90%)
30.000
No offset
0.000
0.000
0.000
0.000
0.000
Sn=xxxx followed by ENTER
Pn=xxxx followed by ENTER
S1=-0.002 followed by ENTER
S1=0.001 followed by ENTER
PsinHg
Static offset
1.600
No offset
3.200
S1 (10%)
6.400
S2 (20%)
9.600
S3 (30%)
12.800
S4 (40%)
16.000
S5 (50%)
16.000
16.000
16.000
16.000
16.000
16.000
16.000
16.000
16.000
16.000
19.200
S6 (60%)
22.400
S7 (70%)
25.600
S8 (80%)
28.800
S9 (90%)
32.000
No offset
Page 68