Section 2 : Controls & Connections; Main Unit, Top Panel - Laversab 6300 User Manual

High accuracy automated pressure controller
Hide thumbs Also See for 6300:
Table of Contents

Advertisement

6300 Rev H User's Manual
SECTION 2 : CONTROLS & CONNECTIONS
2.1

MAIN UNIT, TOP PANEL

The model 6300 top panel provides easy access to all the connections. Please refer to Figure 2.1. (shown on
next page)
[1]
AC INPUT connector :
This is a 3-pin male circular connector. A power cord is provided with the 6300. The circular connector end
of the power cord needs to be connected here. The power requirement of the 6300 is 90-260 VAC, 47-440 Hz
with a maximum power consumption of 100 VA (200 VA with heaters). The pin-out of this connector is
provided in Appendix C
Caution:
Connecting incorrect power to the 6300 will cause considerable damage
[2]
AC Fuse :
A 5x20 mm fuse is located inside the fuse holder. This time-delay fuse, with a rating of 2.0 amps, 250 Volts,
limits AC power to the unit.
[3]
AC On/Off switch :
This toggle switch connects (or disconnects) AC power between the AC INPUT connector and the 6300.
Even when this switch is ON, the 6300 becomes fully operational only after the Power key located on the
keypad of the Remote Unit is pressed for a minimum of 3 seconds.
When an internal battery is included, the battery-charging circuits inside the 6300 become active when the
AC On/Off switch is turned ON.
When internal heaters are included, the heater-control circuits inside the 6300 become active when the AC
On/Off switch is turned ON.
[4]
Battery Fuse :
This fuse is installed only if an internal battery is included. A 5x20 mm fuse is located inside the fuse holder.
This time-delay fuse, with a rating of 10 amps, 250 Volts, limits Battery power to the unit.
125-9106A