6300 Rev H User's Manual
Command
Cm
Read the value for calibration point m, where m is
defined in the table below.
m
1
2
3
4
Cm=n
Set the calibration point m to the value n. This
command required the calibration password to be
set.
CC=n
Calibration complete (n=1)
Calibration Void (n=0)
After this command is sent, the calibration
password is reset. The calibration password is
also reset on power-up
PZ
Read the zero-offset value for the Pitot transducer
PZ=n
This requires the calibration password to be set.
Set the zero-offset value for the Pitot transducer to
n
PS
Read the slope value for the Pitot transducer
PS=n
This requires the calibration password to be set.
Set the slope value for the Pitot transducer to n
SZ
Read the zero-offset value for the Static
transducer
SZ=n
This requires the calibration password to be set.
Set the zero-offset value for the Static transducer
to n
SS
Read the slope value for the Static transducer
SS=n
This requires the calibration password to be set.
Set the slope value for the Static transducer to n
HO
Read the value of 'Hours of Operation'
LC
Read the 'Last calibration date'. The string
returned is in the format 'mm/dd/yy'
LC=n
This command requires the calibration password
to be set. Set the 'Last calibration date' to n,
where n has the format 'mmddyy' (no '/' in
between)
NC
Read the 'Next calibration due' date. The string
returned shows 'mm/dd/yy'.
OC=n
Clears all offset values to zero for Static offsets
(n=0) or Pitot (n=1). This command requires the
calibration password to be set.
125-9106A
Description
Calibration point
Pitot vacuum point
Pitot full-scale point
Static vacuum point
Static full-scale point
Data sent from
Data sent from
PC
None
Value for
calibration point
m
n
0 if OK; or
Error code
n
0 if OK; or
Error code
None
Pitot zero-offset
value
n
0 if OK; or
Error code
None
Pitot slope value
n
0 if OK; or
Error code
None
Static zero-offset
value
n
0 if OK; or
Error code
None
Static slope
value
n
0 if OK; or
Error code
None
Value of Hours
of operation
None
Date as 8-byte
string
n
0 if OK; or
Error code
None
Date as 8-byte
string
n
0 if OK; or
Error code
6300