Laversab 6300 User Manual page 79

High accuracy automated pressure controller
Hide thumbs Also See for 6300:
Table of Contents

Advertisement

6300 Rev H User's Manual
Command
Pm
Read the value for Pitot offset m, where m is as
defined in the table below.
Pm=n
Set the Pitot offset m to a value of n. This
command requires the calibration password to be
set.
SE
Returns details of 13 Self-test parameters in the
following format: parameter code=n, where n=0
(pass) or 1 (fail). The parameter codes are
explained in the table below.
Param.
code
A
B
C
D
E
F
G
H
I
J
K
L
M
125-9106A
Description
m
Pitot offset
1
10% offset
2
20% offset
3
30% offset
4
40% offset
5
50% offset
6
60% offset
7
70% offset
8
80% offset
9
90% offset
Description
Ps control solenoid
Ps output solenoid
Pt control solenoid
Pt output solenoid
Pressure pump
Vacuum pump
Ps control system
Pt control system
Ps main Xd measurement
Pt main Xd measurement
Ps control Xd measurement
Pt control Xd measurement
Baro Xd measurement
Data sent from
Data sent from
PC
None
Pitot offset value
n
0 if OK; or
Error code
None
13 parameters
with pass/fail
code for each
6300
Page 79

Advertisement

Table of Contents
loading

Table of Contents