Laversab 6300 User Manual page 70

High accuracy automated pressure controller
Hide thumbs Also See for 6300:
Table of Contents

Advertisement

6300 Rev H User's Manual
Model: 6300
Full scale:
Last Calibrated date:
___________________________________________________________________________
TEST POINT
1. Vacuum
2. Fullscale
TEST POINT
1. Vacuum
2. Fullscale
___________________________________________________________________________
STATIC PRESSURE (absolute)
APPLIED
inHg
1.600
3.200
6.400
9.600
12.800
16.000
19.200
22.400
25.600
28.800
32.000
___________________________________________________________________________
Calibrated by:
125-9106A
Static: 32 inHg
Pitot: 30 QcinHg (differential)
08/15/14
CALIBRATION RESULTS
STATIC OUTPUT
AS FOUND
inHg
1.601
32.002
PITOT OUTPUT
AS FOUND
QcinHg
0.002
30.003
VERIFICATION RESULTS
DISPLAYED
inHg
1.000
3.200
6.401
9.600
12.800
16.000
19.200
22.401
25.600
28.800
32.000
Figure 6.10 Sample calibration report
Serial#: 82678
AS LEFT
inHg
1.600
32.000
AS LEFT
QcinHg
0.000
30.000
PITOT PRESSURE (differential)
APPLIED
QcinHg
0.000
3.000
6.000
9.000
12.000
15.000
18.000
21.000
24.000
27.000
30.000
Date:
DEVIATION
inHg
0.001
0.002
DEVIATION
QcinHg
0.002
0.003
DISPLAYED
QcinHg
0.000
2.999
6.000
9.000
12.000
15.000
18.001
21.000
24.000
27.000
30.000
Page 70