6300 Rev H User's Manual
At the end of two minutes, the screen will appear similar to Figure 3.23 below.
At the end of three minutes, the screen will appear similar to Figure 3.24 below.
At the end of three minutes, the screen will appear similar to Figure 3.24 below.
The parameters indicated in Figures 3.21 through 3.24 are explained below.
The parameters indicated in Figures 3.21 through
[1] Leak Timer: Shows the elapsed time since the Leak-check was started.
[1] Leak Timer: Shows the elapsed time since the Leak
[2] Start value: The Actual value at the start of the leak
[2] Start value: The Actual value at the start of the leak-check (at time 0:00)
[3] Leak timer values: Currently set at 1, 2 and 3 minutes. Can be changed to values up to 20 minutes.
[3] Leak timer values: Currently set at 1, 2 and 3 minutes. Can be
See Section 3.10.7 for details.
[4] The Actual value at the end of 1 minute
[4] The Actual value at the end of 1 minute
[5] Total leak in 1 minute
[6] Average per-minute leak rate over 1 minute
minute leak rate over 1 minute
125-9106A
Figure 3.22 : Pitot Leak (1-min)
screen will appear similar to Figure 3.23 below.
Figure 3.23 : Pitot Leak (2-min)
Figure 3.24 : Pitot Leak (3-min)
changed to values up to 20 minutes.
Page 36