6300 Rev H User's Manual
How the user can change the Static Target value is described in Section 3.4.
[11]
Static-Rate units
The Static-Rate units are typically the Static units per minute. If Static units are Feet then the rate units
are Feet/min. However, there is one exception as shown in the table below. Static-Rate units cannot be
changed by the user.
[12]
Static-Rate Actual value
The Actual rate of change of pressure at the Static output, converted to current Static-Rate units, is
displayed here. The displayed resolution depends on the Static-Rate units as shown in the table below.
The Actual Rate value is updated every 250 msec.
[13]
Static-Rate Target value
The rate at which the Static Target value ramps from one target to the next target, is determined by the
Static-Rate Target value. When units are Feet or Meters, this rate is the Climb-rate or Vertical Speed.
The Static-Rate target value has the same resolution as the Static-Rate Actual value. How to set the Static-
Rate Target is described in Section 3.5.
125-9106A
Exception
Static units
Static-Rate units
Meters
Meters/sec
Static-Rate
Units
Feet/min
1
Meters/sec
0.01
inHg/min
0.001
mbar/min
0.01
psi/min
0.001
mmHg/min
0.01
kpa/min
0.001
Displayed
Resolution
Page 20