Laversab 6300 User Manual page 30

High accuracy automated pressure controller
Hide thumbs Also See for 6300:
Table of Contents

Advertisement

6300 Rev H User's Manual
Static Rate Target value
The Static Rate Target value can be changed only if the key# symbol
(7)
is displayed to the right of the
Static Rate Target value field. Press key '7' on the keypad and the Lower section of the display will
appear as shown in Figure 3.13 below.
Figure 3.13 : Static Rate Target value
Use the keypad to enter the desired value in the field shown. To clear an entry, press CLR. To accept the
entry, press ENTER. To exit this screen without changing the Rate Target value, press CANC. If the
number of digits entered (either before or after the decimal) exceeds the maximum allowed for the
current units, a long beep will be heard.
If the entered value exceeds the programmed limits, the following message will appear.
Static Rate target exceeds limit. Press
CLR
Note: When Static is in pressure units, the commanded Static Rate target is limited by the pressure-
equivalent of the Static Rate target that has been set for altitude units (feet/min or meters/sec). This
ensures that the rate of change of Static pressure does not exceed the target altitude rate.
Note: In pressure units, Static-Rate targets can be set between 0 and 30 inHg/min.
If the Rate Target value needs to be changed in small increments, instead of entering a value, the arrow
keys may be used to Jog the Rate Target value up or down. If the current Static Rate Target value is 2000
Feet/min, and the Lower screen appears as in Figure 3.13 above, pressing the
Up-Arrow
key will make
the screen appear as shown in Figure 3.14 below.
125-9106A
Page 30