Note: Pt = Pitot and Ps = Static
2.1

MAIN UNIT TOP PANEL

The model 6300-M3 top panel (Figure 2.1) provides easy access to all the connections.
[1]
AC INPUT connector :
This is a 3-pin male circular connector. An AC power cord is provided with the 6300-M3. The
circular connector end of the power cord needs to connected here. The power requirement of
the 6300-M3 is 90-260 VAC, 47-440 Hz with a maximum power consumption of 300 VA.
Caution: Connecting incorrect power to the 6300-M3 will cause considerable damage
The power cord is normally left connected to the 6300-M3 at the AC INPUT connector and is
wrapped around two cord posts. The cord retainers on top of these posts swivel in either
direction and latch at 180 degree positions, allowing easy wrapping and unwrapping of the
power cord. A cord clamp allows the cord to be held in place after it is wrapped.
[2]
AC Fuse :
A 5x20 mm fuse is located inside the fuse holder. The fuse is a time-delay fuse with a rating of
3.15 amps, 250 Volts.
[3]
AC On/Off switch :
This toggle switch connects (or disconnects) AC power between the AC INPUT connector and
the 6300-M3. Even when this switch is ON, the 6300-M3 becomes operational only after the
On/Off switch located on the right side of the Remote Unit is turned ON. (Push in to turn
ON)
[4]
DC INPUT connector:
This is a 4-pin male circular connector. A DC power cord is provided with the 6300-M3. The
DC power requirement of the 6300-M3 is 16-36 VDC (nominal 28 VDC), with a maximum
power consumption of 300 W.
Caution: Connecting incorrect DC power to the 6300-M3 will cause considerable damage
SECTION 2

CONTROLS AND CONNECTIONS

2