10. Loosely attach the switch mounting bracket to the eyewash mounting brackets with four
0.75 hex head screws, four
11. Loosely attach the 10 Series GO Switch to the switch mounting bracket with the hardware provided.
12. Attach the target magnet to the target arm using two #10–24 x 0.75 flat head screws, two #10 lock
washers, and two #10–24 hex nuts.
13. Place the target arm on the face of the eyewash handle and adjust the position of the mounting
brackets, 10 Series GO Switch, and target arm so that the target magnet properly engages the
switch. Mark the position of the slot of the target arm. Drill two holes in the eyewash handle as
shown to be able to attach the target arm.
14. Attach the target arm to the eyewash handle using two #10–24 x 0.50 round head screws, two #10
lock washers, and two #10–24 hex nuts.
15. Tighten hardware to hold the mounting brackets, 10 Series GO Switch, and target arm in place.
16. Set up and calibrate the GO Switches as described in the 10 Series GO Switch operation and
installation instructions.
17. Make supply and electrical connections to the GO Switches as described in the 10 Series GO Switch
operation and installation instructions.
1
Safety Shower Monitoring
/
" plain washers, four 1/4" lock washers, and four
1
4
2
ITEM
QTY
1
1
KIT, SHOWER MOUNTING
2
1
KIT, EYEWASH MOUNTING
Parts List
DESCRIPTION
Safety Shower Monitoring
/
1
/
–20 hex nuts.
1
4
1
2
July 2017
–20 x
4
124