Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers
Appendix A – Reagents
A.1 Reagent Solutions
Warning.
Ensure personal protective equipment (PPE) such as
gloves and eye protection are worn during any
maintenance and that any spillages are cleaned-up.
Observe all health and safety procedures for handling
chemicals.
A standard set of reagents for an Aztec 600 analyzer (except an
Aztec 600 color analyzer) comprises 3 reagents (5 l of each) and
a high calibration standard (2.5 l). The Aztec 600 color analyzer
does not use reagents for measurement; a high calibration
standard is used only for automatic calibration.
Because the usage of de-ionised water depends on the
measurement range (see Table A.1) this is supplied separately
and available in 10 or 25 l sizes.
Aztec 600
Reagent 1
Reagent 2
Analyzer
(5 l)
Aluminium
Acid
Ammonia
Color
Color
N/A
Iron
Acid
Manganese
Buffer
Manganese
Preconditioner
Low Range
Acid
Phosphate
Molybdate
Reagent
Table A.1 Reagent Sets/Standard Measurement Range
The default calibration standards are selected to give accurate
results across the most frequently used measurement ranges.
IM/AZT6CR–EN Rev. J
Reagent 3
Standard
(5 l)
(5 l)
(2.5 l)
0.2
Buffer
Color
ppm Al
0.35
Alkaline
ppm NH
Catalyst
Reagent
(0.288 ppm
NH
50 Hazen
N/A
N/A
(50 mg/l
Pt-Co)
0.2
Buffer
Color
ppm Fe
Complexing
0.5
Color
Agent
ppm Mn
Buffer/
0.1
Oxidizing
Color
ppm Mn
Agent
1 ppm PO
Reducing
N/A
(3.066 ppm
Agent
PO
A.1.1 Reagent Storage
Store reagents in their original packaging, in a cool, dry, well
ventilated environment.
Note.
Do not use reagents outside their shelf life.
The following reagents are sensitive to light and must
be stored in dark bottles:
Ammonia – Reagents 2 and 3
Low Range Manganese – Reagent 3
Contact with visible, ultraviolet or infrared radiation
shortens the shelf life of these reagents.
2 types of reagent bottle kit are available from ABB – see
Appendix F.3, page 96.
A.2 Reagent Consumption
The reagent usage depends on how many samples an hour are
measured.
Typically 0.75 ml of each reagent is used per determination.
Samples Measured Per Hour
3
1
-N)
3
2
3
4
5
6
Table A.2 Reagent Consumption
-P
4
)
4
Appendix A – Reagents
Duration of Reagent Set
(Days)
240
120
80
60
50
40
77

Hide quick links: