Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers
10.2 Cell Diagnostics
Cell Diagnostics enables the user to control the operation of the analyzer manually – bringing in reagents and sample etc. It also
displays the detector output in real time giving an insight into how the analyzer is performing.
Note.
If this procedure is selected when the analyzer is in
operation, a warning is displayed:
Fields
Operation
Volume
Port
Detector Voltage
LED Current
10.3 Relay Test
All of the alarm relays can be set individually or reset to check their operation.
Note. This procedure affects the analyzer relay functions and the analyzer
10.4 Current Output Test
Within the Current Output Test Screen the user is able to check the analyzer current outputs manually.
A calibrated ammeter is required for this test. Put the leads from the ammeter onto the two current output terminals on the I/O board
– see Section 4.6.3, page 18.
Check that the value shown on the analyzer LCD is the same value shown on the ammeter.
Use the
and
keys to increase/decrease the milliamp value.
If the ammeter and analyzer do not agree it may be necessary to recalibrate the current outputs – see Section 6.5.2, page 36.
Note. This procedure affects the current outputs of the analyzer and the analyzer
54
Description
Enables the user to bring in solutions, expel solutions or reset the pump.
Enables the user to specify the quantity of liquid (in ml).
Enables the user to specify the port (and therefore liquid) to import. Only one port can be opened at
any one time.
Displays the detector voltage (0 to 4095 mV) in real time.
Displays the LED current in real time.
Note.
Select
and press
to return to normal operation. Select
Back
and press
to continue to cell diagnostics. This
Continue
stops the current analyzer operation:
the normal state.
overrides
their normal state
overrides
10 Diagnostics
IM/AZT6CR–EN Rev. J