Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers

6.8 Commissioning

Note. Access to the Commissioning Level is prevented by an Internal Security Switch. The switch is, by default, set to 'Disabled'
and should be set to 'Enabled' only if the measurement parameter or number of streams is to be changed. For further information
on the function of the Internal Security Switch, refer to Section 6.1.4, page 28.
There is one tab in the Commissioning screen:

6.8.1 Setup

Fields
Instrument type
Streams Fitted
Serial Number
Mode of Operation
Clear NonVol
Dilution Point
IM/AZT6CR–EN Rev. J
Description
The parameter to be measured by the analyzer. The options are:
Aluminium Monitor
Ammonia Monitor
Iron Monitor
Manganese Monitor
Manganese LR Monitor
Phosphate Monitor
Colour Monitor
The number of streams to be measured by the analyzer; 1, 2 or 3.
The serial number of the analyzer.
Options are:
Normal Run Mode
Demonstration Mode
Test Mode
Select 'Yes' to clear the analyzer's non-volatile memory.
6 Configuration
41