Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers

4.6.1 Access to Connections

Note.
Cable entry holes are located on both sides of the enclosure.
Application board connection labels for the terminal blocks are identified in Fig. 4.7 on page 18.
Fig. 4.5 Accessing and Making Electrical Connections
Referring to Fig. 4.5:
1. Turn the two electronics section door retaining screws
1
/
turn counter-clockwise and open the door.
4
2. Using a cross-head screwdriver, remove the four screws
B
and remove the transparent cover plate
3. For each cable entry, slide retaining clip
E
plug
and remove the blanking plug.
F
4. Fit cable gland
and secure using nut
H
5. Remove gland cover
gland cover.
6. Route the cable through cable gland
enclosure case.
Note. Cable glands are supplied with single- and
twin-holed bushes. Use the single-holed bush for the
mains power cable.
16
A
B
C
.
D
off blanking
G
.
I
and route cable
through the
F
and through the
D
C
J
G
7. Remove each terminal block connection plug
using a small flat-bladed screwdriver, make connections
A
to each plug. Ensure wires are connected to the correct
terminals – see Fig. 4.7, page 18.
8. Reconnect the terminal block plugs to the appropriate
sockets on the application board.
9. Tighten gland nut
10. If required, connect the Ethernet cable – see Section
4.6.2, page 17.
11. When all the connections are made, replace the
transparent cover plate
B
screws
. Close the door to the electronics section and
secure by turning the two door retaining screws
1
/
turn clockwise.
4
.
4 Installation
E
H
F
H
for each connection made.
C
and secure using the four
IM/AZT6CR–EN Rev. J
I
J
and,
A