Completion - ABB Aztec 600 User Manual

Aluminium, ammonia, color, iron, manganese, phosphate single- and multi-stream colorimetric analyzers
Hide thumbs Also See for Aztec 600:
Table of Contents

Advertisement

Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers
9. Referring to Fig. 13.14, refit the motor and sample tube
O
to the lower half of the measurement cell, ensuring
that the motor wiring plug is aligned with the socket (see
M
in Fig. 13.13). Secure the motor mounting plate to the
lower half of the measurement cell with the four screws
N
.
10. Referring to Fig. 13.13, reconnect the motor wiring plug
M
.
11. Referring to Fig. 13.12, refit the measurement cell top
L
cover
and secure with the two screws
12. Referring to Fig. 13.11,
a. Fit a new O-ring
located correctly in the groove in the end cap.
b. Invert the measurement cell assembly and position
end cap H over the end of the sample tube
c. Refit base plate
centered in the hole in the base plate. Secure the
base plate with the four screws
13. Referring to Fig. 13.10, refit cover
14. Referring to Fig. 13.9:
a. Locate the measurement cell assembly onto the
analytical section back plate and secure with the
C
four M4 screws
b. Reconnect measurement cell ribbon cable
IM/AZT6CR–EN Rev. J
K
.
I
H
to end cap
, ensuring it is
G
ensuring that the end cap is
F
.
E
.
.
B

13.3.7 Completion

1. Referring to Fig. 13.8, raise the analytical section to the
upright position and secure with handle
2. Referring to Fig. 13.6, connect the new interconnect tube
to the connector
assembly.
3. Turn on the sample feed to the side sample pot.
4. Restore the power supply to the analyzer and select
& Calibrate
page 51.
J
.
.
A
C
at the base of the measurement cell
from the operator menu – see Section 9.4,