Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers
Mechanical Data
Ingress protection
IP31*
Sample connections
Inlet:
6 mm OD push-fit x
Outlet:
10 mm OD push-fit x
Dimensions
Height
653 mm (25.7 in)
Width
366 mm (14.4 in) max.
Depth
183 mm (7.2 in) door closed
430 mm (16.9) door open
Weight
15 kg (33 lb)
Materials of construction
Electronics enclosure
10 % glass loaded polycarbonate
Main enclosure
Noryl
Lower tray
20 % glass loaded polypropylene
Door
Acrylic
Electrical
Supply ranges
100 to 240 V max. AC 50/60 Hz ± 10 % (90 to 264 V AC, 45/65 Hz)
18 to 36 V DC (optional)
Power consumption
75 W max. – AC
100 W max. – DC
Analog Outputs
Single and multi-stream analyzers
6 isolated current outputs, fully assignable and programmable over a
0 to 20 mA range (up to 22 mA if required)
Alarms/Relay Outputs
Single- and multi-stream analyzers
One per unit:
 Stop relay
 Attention relay
 Failure relay
 Calibrate relay
Six per unit:
 Fully user-assignable alarm relays
Rating
Voltage
250 V AC
Current
5 A AC
Loading (non-inductive)
1250 VA
*Not evaluated for UL or CB
76
Connectivity/Communications
Ethernet connection
Bus communications
1
/
in BSP elbow
4
3
/
in BSP elbow
8
Data Handling, Storage and Display
Security
Storage
Trend analysis
Data transfer
Approvals, Certification and Safety
Safety Approval
CE Mark
General Safety
EMC
Emissions & immunity
30 V DC
5 A DC
150 W
Web server with ftp
For real-time monitoring, configuration,
data file access and e-mail capability
Profibus DP V1.0 (optional)
Multi level security
Operator and configuration
Password or security switch
Removable Secure Digital (SD) card
Local and remote
SD card or FTP
cULus
Covers EMC & LV Directives (including latest version EN 61010)
EN 61010–1
Overvoltage Class 11 on inputs and outputs
Pollution category 2
Meets requirements of IEC61326 for an Industrial Environment
14 Specification
DS/AZT6AL–EN Rev. F
DS/AZT6AM–EN Rev. C
DS/AZT6C–EN Rev. B
DS/AZT6IR–EN Rev. H
DS/AZT6ML–EN Rev. C
DS/AZT6MN–EN Rev. D
DS/AZT6P–EN Rev. D