Appendix E - Updating The Software - ABB Aztec 600 User Manual

Aluminium, ammonia, color, iron, manganese, phosphate single- and multi-stream colorimetric analyzers
Hide thumbs Also See for Aztec 600:
Table of Contents

Advertisement

Aztec 600 aluminium, ammonia, color, iron, manganese, phosphate
Single- and multi-stream colorimetric analyzers
Appendix E – Updating the Software
Normal
Operation (Up)
Software
Update
(Down)
B
A
Fig. E.1 SD Card Removal
1. Stop analyzer operation by selecting 'Operate/Stop' from
the main operator menu.
2. Ensure the analyzer is switched off.
3. Use a large flat-headed screwdriver to release the two
A
door catches
.
B
4. Remove the SD card
5. Refer
to
the
documentation
(http://www.abb.com)
6. Download the software and transfer it to a blank SD card
using an appropriate card reader.
C
7. Set the switch
(above the SD card socket) to the lower
position (Software Update).
8. Insert the SD card, close the door and switch on the
analyzer. A start-up message appears for a short duration
(approximately 5 seconds) followed by a 'File Load
Progress' bar then the 'ABB' splash screen:
94
C
.
on
the
web
for details of any software updates.
9. A 'Warning' screen appears next. This provides the option
to upgrade the existing system software or exit without
upgrading:
10. To proceed, press the
11. The 'Commissioning V2.0' screen appears with the
'Upgrade Firmware' button selected:
site
12. To upgrade the existing system software, press the
key to start the process. The software is transferred to the
analyzer in sections.
13. Press the
may take a few minutes).
14. Open the door and move the switch to the up position.
15. Close the door and re-tighten the door catches (
E.1).
16. Turn power off and then on to restart the analyzer.
Appendix E – Updating the Software
key.
key as each section is installed (installation
IM/AZT6CR–EN Rev. J