6.2.3 Conduct a Factory standard
calibration
–
Press the ON / OFF key
=>
The standard menu will appear on
the screen
–
Press the u key to select the System
menu
Select the submenu Customer
–
calibration with the q key
–
Press 4 to confirm
–
Press the q key to select the Factory
calibration
–
Press 4 to confirm or press Ó to
cancel selection
–
Enter the value for Calibration Min
(e.g. - 1.000 mm). Press the u key
to select the required position (digit).
Press the q p keys to enter the
required value.
–
Press 4 to confirm or press Ó to
cancel selection
Mahr GmbH • Millimar C1200
Measurement
Display
Setup
Absolute mode
Customer calibration
Factory calibration
Factory Settings
Info
6
3
Measurement
Display
Setup
Customer calibration
Enable
Probe calibration
Factory standard calibration
Customer calibration Reset
6
3
Measurement
Display
Setup
Customer calibration
0 0 1 . 0 0 0
-
Calibration Min
Nominal value
-1.000
+0
u
3
6
System
ü
q
p
System
q
p
System
Calibration Max
1.000
q
p
37