Summary of Contents for Mahr Millimar S 1840

  • Page 1 Betriebsanleitung Operating instructions Guide de l’utilisateur Manuale di istruzioni Millimar S 1840 Mahr GmbH · Carl-Mahr-Straße 1 · D-37073 Göttingen · Telefon +49 551 7073-0 · Fax +49 551 71021...
  • Page 2: Permitted Uses

    Congratulations on choosing a product by Mahr Permitted Uses GmbH. We kindly request that you follow the The Millimar S 1840 measuring instrument is easy instructions below in order to ensure the long- to operate and designed for use in production.
  • Page 3: Disposal

    The crossed-out garbage can symbol indicates that separate collection is necessary. Mahr GmbH carries out the redemption and disposal of its electrical and electronic products in accordance with legal requirements. Please contact your local service representative.
  • Page 4: Safety Instructions

    12. Before opening the housing, disconnect the power supply. 13. The gages with which the Millimar is used are subject to gage monitoring. For this reason, gage monitoring performed by the user or Mahr Service must ensure adherence to the specified error limits.
  • Page 5 Mahr GmbH, Millimar S1840...
  • Page 6: Table Of Contents

    ..... 42 Assigning frequently used functions Performing a two point master onto the MENU key ........27 measurement ............ 43 Positioning the probe in the Original master measurement ....43 measuring device (Setup) ....28 Mahr GmbH, Millimar S1840...
  • Page 7 14.7 Connecting to the PC using a zero modem cable (basic circuit diagram) . 75 12.1 Performing a compound (sum) measurement ............ 60 12.2 Performing a difference measurement.. 12.3 Performing a radial runout measurement .............61 Measuring value memories ....62 Mahr GmbH, Millimar S1840...
  • Page 8 15.4.2 Setting the sensitivity of the analog output ........83 Restoring factory settings ....84 Error messages ........85 Technical data ........86 Contacts at Mahr ........88 Warranty ..........88 Index ............89 Navigating in the catalog of functions and parameters ....94 Mahr GmbH, Millimar S1840...
  • Page 9: Commissioning

    Damage that occurs 6. Press START as soon as the required language due to shipping in unsuitable packaging is displayed. is not part of the Mahr GmbH warranty The flashing entry "MM" appears on the obligation! display. 7. Use the DATA and MASTER keys to select 2.
  • Page 10 Use the ESC key to return to the higher function the catalog of functions and parameters. level. Use the START key to accept the parameter settings. This automatically returns the user to the higher function level. Mahr GmbH, Millimar S1840...
  • Page 11 MENU key. Windows software "Millimar D 1000 S" The numerical value is then displayed with the is available on the Mahr GmbH website preceding sign flashing. ("www.mahr.de") (web code #6776). – Use the DATA and MASTER keys to set the The most important settings can be preceding sign ("+"...
  • Page 12: General

    General The Millimar S 1840 measuring instrument is easy Depending on the interface protocol selected, to operate and designed for use in production. the RS-232 interface enables the transfer/ Two versions are available: query of measuring results to/from an external...
  • Page 13: Operating Elements And Connections

    Millimar's catalog of functions and parameters. S 1840-M S 1840 PE S 1840-F Fig. 1 Front panel of the Millimar S 1840 1 Light bar 2 Display 3 Keypad 4 Connection for pneumatic measuring equipment Mahr GmbH, Millimar S1840...
  • Page 14 Press START again to accept the numerical value. If the DATA key is pressed after a measurement is completed, data are sent to a connected PC if the interface is set accordingly (see section 14). Mahr GmbH, Millimar S1840...
  • Page 15 "MM (4.4. 1 )" is displayed. — Use the DATA and MASTER keys to set the unit of measurement and accept it by pressing the START key twice. The message "MAHR" is displayed and then the current probe value. Mahr GmbH, Millimar S1840...
  • Page 16 PROCESS (of measurement), pressing START either starts individual measurements or measuring cycles. If the approproate settings have been set, START is used not only to start measurements, but also to end them (see section 10.2). Mahr GmbH, Millimar S1840...
  • Page 17: Rear Panel Of The Millimar

    S 1840-M S 1840 PE S 1840-F Fig. 2 Rear panel of the Millimar S 1840 1 ON/OFF switch 2 Connection for AC adapter 3 Interface for additional instruments (25-pin D-Sub) 4 Serial RS-232 interface...
  • Page 18 Millimar series (e.g. foot switch) or a PLC The label is designed for original Mahr probes in for automatic sorting procedures (see section 15, the P20xx series.
  • Page 19: Making Basic Settings

    6. Use the DATA and MASTER keys to select functions and parameters and make the the new display language (Deutsch, English, required settings. Franç., Ital., Español, Portug. or Svenska). 7. Press START. The designation of the selected language flashes. Mahr GmbH, Millimar S1840...
  • Page 20: Setting Light Bar Brightness

    7. Use the DATA and MASTER keys to select required settings. the new brightness level: (HIGH, MIDDLE, LOW). 8. Press START. The designation of the selected brightness level flashes. Mahr GmbH, Millimar S1840...
  • Page 21: Setting The Unit Of Measurement

    7. Press START. The designation of the selected If further settings are required, use the ESC, unit of measurement flashes. MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 22: Setting The Resolution/Display Format Of The Measuring Value

    7. Use the DATA and MASTER keys to select functions and parameters and make the the new display format. The following can be required settings. selected: µm inch ooo.oo ooooo ooo.oooo ooo.ooo ooooo.o o.ooooo ooo.oooo ooooo.oo* o.ooooo5 o.ooooo o.oooooo Mahr GmbH, Millimar S1840...
  • Page 23: Selecting The Number Of Features/Test Results To Be Displayed

    The display switches from one feature to the other when the specified section limits are exceeded or when a control signal is transmitted via the I/O interface. Mahr GmbH, Millimar S1840...
  • Page 24 In the top line of the required settings. display, instead of the currently valid display range, the numerical value of the first feature is shown; the value of the second feature is shown in the bottom line. Mahr GmbH, Millimar S1840...
  • Page 25: Setting The Light Bar Display Range

    These display ranges are only available if the maximum resolution was set under SETTING->DISPLAY- >RESOL.". In this case, the measuring ranges is limited to ± 0.2 mm. Mahr GmbH, Millimar S1840...
  • Page 26 If further settings are required, use the ESC, MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 27: Setting The Origin Of The Display Range

    8. Press START. The selected origin flashes. If further settings are required, use the ESC, MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 28: Assigning Frequently Used Functions Onto The Menu Key

    MENU key. again. Confirm the prompt with START. The deletion of the M E N U key assignments can be canceled with ESC. Mahr GmbH, Millimar S1840...
  • Page 29: Positioning The Probe In The Measuring Device (Setup)

    DATA or MASTER. The next smaller or larger display range will then be employed. Press either of the keys again in order to increase or decrease the size of the display range according to the programmed gradations. Mahr GmbH, Millimar S1840...
  • Page 30: Calibrating To Compensate For Probe Sensitivity

    C1 can be determined (calibrated) then the sensitivity of the probe connected to C2 adjusted to that of the first probe. Mahr GmbH, Millimar S1840...
  • Page 31: Calibrating A Measuring Channel

    START. 10. Set the nominal value adjuster to the negative calibration value and confirm with START. „POS VAL (4.9. 1 . 1 . 1 . 1 ) is displayed. Mahr GmbH, Millimar S1840...
  • Page 32: Adjusting The Sensitivity Of The Probe At C2 With Respect To That Of The Probe At C1 (Instruments With Inductive Probes Only)

    START. "CHAN 1 (4.9. 1 ) is displayed. to the corresponding item in the catalog of 8. Use MASTER to set "CHAN 2 (4.9.2)". functions and parameters and make the required settings. 9. Use the MENU key to select the setting "OFFS LO". Mahr GmbH, Millimar S1840...
  • Page 33: Making Settings For Calculating And Displaying Features

    Setting the factor for correcting the pneumatic or up to two inductive probes can be displayed value of the feature connected to the Millimar S 1840. One or two features are calculated from the raw value(s) (see Certain measuring tasks may require the entry of the section "Selecting the number of features/...
  • Page 34 START key. The numerical input closes and required settings. the value set is flashing. If the factor of a linking formula is often changed, this catalog function can be assigned to the MENU key to enable rapid access. Mahr GmbH, Millimar S1840...
  • Page 35: Selecting A Channel Or Channel Linking

    C2 – C1 Difference of the raw values of the probes at C1 and C2 If C2>C1, the difference is positive; if C2 < C1, it is negative -C1 -C2 Difference between the two inverted raw values of the probes at C1 and C2 Mahr GmbH, Millimar S1840...
  • Page 36 If the linking formula (channel linking) is often changed, this function can be assigned to the MENU key to enable rapid access. Mahr GmbH, Millimar S1840...
  • Page 37: Selecting Features

    (minimum value) is always + Measuring value 2 + ..shown. The value displayed + Measuring value x) ÷ only changes if a value falls Number x of single measuring below the currently valid values minimum value. Mahr GmbH, Millimar S1840...
  • Page 38 If the function (feature) is often changed, this function can be assigned to the MENU key to enable rapid access. Mahr GmbH, Millimar S1840...
  • Page 39: Master Measurement

    "FRM FEH" appears. before the measuring result is displayed. The characteristic curve of the probe thus undergoes a parallel shift in the display range (see Fig. 3). Mahr GmbH, Millimar S1840...
  • Page 40 10 Actual value of MSTR.MAX taking the linear error into consideration 11 Nominal value of MSTR.MAX 12 Actual pitch of the characteristic curve of the probe 13 Required pitch of the characteristic curve of the probe to display correct values Mahr GmbH, Millimar S1840...
  • Page 41: Selecting The Type Of Master Measurement

    START. The current measuring value or result is displayed again. If further settings are required, use the ESC, MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 42: Performing A One Point Master Measurement

    If further settings are required, use the ESC, MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 43: Entering The Nominal Value For A Two Point Master Measurement

    START key returns you to the to the corresponding item in the catalog of numerical input (see point 4). The nominal functions and parameters and make the value displayed can be changed again, if required settings. necessary. Mahr GmbH, Millimar S1840...
  • Page 44: Performing A Two Point Master

    For the two point measurement, the change in sensitivity (= change in pitch) is not monitored. . The original master measurement is deleted by selecting the "PROCESS -> CLR" menu. The next master measurement then automatically becomes the original master measurement. Mahr GmbH, Millimar S1840...
  • Page 45: Setting Tolerances And Limit Values

    (see point 7). The 6. Press the MENU key. The current value of the numerical value displayed can be changed upper tolerance limit is displayed below the again. "TOLER. +" entry with a flashing preceding sign. Mahr GmbH, Millimar S1840...
  • Page 46 13. When the preceding sign and the numerical value have been set as required, press the START key. The numerical input closes and the value set is flashing. Mahr GmbH, Millimar S1840...
  • Page 47: Setting Warning Limits

    "ALARM-" entry with a flashing preceding 7. Change the preceding sign and numerical sign. value of the upper warning limit as required (see the section "Changing numerical values for parameter settings" in section 0.) Mahr GmbH, Millimar S1840...
  • Page 48 The catalog item "ALARM- (1.4.5)" is displayed again. Pressing either the MASTER or DATA key instead of the START key returns you to setting the numerical value (see point 11). The numerical value displayed can be changed again. Mahr GmbH, Millimar S1840...
  • Page 49: Setting The Color Of The Warning And Tolerance Limits

    6. Press the MENU key again. The entry (1.4. 1 .3.2)" flashes. "> T RED. + (1.4. 1 . 1 . 1 )“ flashes. 7. If necessary, use MASTER to select the "> T YELW + (1.4. 1 . 1 .2)" setting. Mahr GmbH, Millimar S1840...
  • Page 50 If further settings are required, use the ESC, MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 51: Setting Plausibility Limits

    The upper plausibility limit can also be set with a negative value! However, when setting the lower plausibility limit, please note that it must be lower, i.e. greater in the negative direction, than the upper limit. Mahr GmbH, Millimar S1840...
  • Page 52 0.) If a negative value was set for the upper plausibility limit, when setting the lower plausibility limit, please note that it must be lower, i.e. greater in the negative direction, than the upper limit! Mahr GmbH, Millimar S1840...
  • Page 53: Setting Parameters For Recording Measuring Values

    6. Use the DATA and MASTER keys to set the functions and parameters and make the required settings. required limit value. 7. Press START as soon as the required value is set. The selected value flashes. Mahr GmbH, Millimar S1840...
  • Page 54 Disturbing signals (grooves, dirt) affect the measuring result B Low limiting frequency Disturbing signals are filtered out for the most part; measuring result is not affected C Limiting frequency too low The measured signal is distorted. Mahr GmbH, Millimar S1840...
  • Page 55: Specifying The Measuring Process

    "T-PAUSE" expires. Press START to end the succession of single measurements and pauses in between. The letter "T " shown in the bottom line of the display during the measurement indicates an ongoing measurement. Mahr GmbH, Millimar S1840...
  • Page 56 (3. 1 .3)" is displayed again. required settings. Pressing either the MASTER or DATA key instead of the START key returns you to the numerical input (see point 17). A different pause period can be entered. Mahr GmbH, Millimar S1840...
  • Page 57: Normal" Mode

    PROCESS -> MEASURE-> MEASMOD (3. 1 . 1 )", the calculated sum or difference or the average is continuously updated. Use START to save the current value. A new measurement starts right away. Mahr GmbH, Millimar S1840...
  • Page 58: Entering A Start Delay

    7. Enter the required start delay in seconds (see the section "Changing numerical values for parameter settings" in section 0.) 8. Press START as soon as the start delay time has been entered. The numerical input closes and the set delay is flashing. Mahr GmbH, Millimar S1840...
  • Page 59: Password Protection

    If the password is forgotten, the instrument must be reset to factory settings (see section 16). Mahr GmbH, Millimar S1840...
  • Page 60: Performing Measurements

    Fig. 6 Probe signals and direction of travel Mahr GmbH, Millimar S1840...
  • Page 61: Performing A Compound (Sum)

    4. Remove the setting master and insert the 4. Remove the setting master and insert the workpiece into the measuring device. workpiece into the measuring device. The workpiece thickness is displayed. The height of the step is displayed. Mahr GmbH, Millimar S1840...
  • Page 62: Performing A Radial Runout

    MASTER (1.8.) The measurement is concluded and the MSTR.MAX –> +000.0000 radial runout deviation of the workpiece is MENU –> PROCESS (3.) –> MEASURE (3. 1 .) displayed. -> MEASMOD -> AUTOMAT (3. 1 . 1 .2) Mahr GmbH, Millimar S1840...
  • Page 63: Measuring Value Memories

    Measuring value memories The Millimar S 1840 has one memory each for When measuring a rotating shaft, for example, the maximum value, the minimum value and the various workpiece features can be derived from average. The stored values are calculated from...
  • Page 64 The measuring value memory can also be started and ended using control signals. This appoach is useful if, for example, the measuring value memory is to be activated and deactivated using contacts mounted on the measuring device. Mahr GmbH, Millimar S1840...
  • Page 65: Working With The Serial (Rs-232) Interface

    For this reason, suitable settings must be selected various instruments exchange data with each for the following parameters: other. Because the Millimar S 1840 operates with different instruments (printers, computers, control Interface protocol (PROTOCL): units, ...) there are several interface protocols from...
  • Page 66 (e.g. Hyperterm), a "break" signal can be triggered with the key combination C T R L + PAU S E / B R E A K . Th e "DTR/DSR" line can be deactivated using the pull-down menu "Call--> Disconnect". Mahr GmbH, Millimar S1840...
  • Page 67: Selecting The Data Transfer Format

    Seven data bits and two stop bits to the corresponding item in the catalog of are transferred. A parity check is functions and parameters and make the carried out, which must produce required settings. an even value. Mahr GmbH, Millimar S1840...
  • Page 68: Selecting The Transfer Rate

    START. The current measuring value or result is displayed again. If further settings are required, use the ESC, MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 69: Selecting A Handshake

    START. The current measuring value or result is displayed again. If further settings are required, use the ESC, MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 70: Selecting How The Data Transfer Is To

    START. The current measuring value or result is displayed again. If further settings are required, use the ESC, MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 71: Examples Of Interface Configurations

    FORMAT: HANDSHK: NONE BAUD: 9600 TRANSM: MANUAL or AUTOM. If the "MANUAL (4.6.5. 1 )" setting was selected under "TRANSM (4.6.5)", the data are transferred to the printer by pressing the DATA key on the Millimar. Mahr GmbH, Millimar S1840...
  • Page 72: Optorsd Interface Protocol

    Requesting the manufacturer and instrument CHA-1-2<CR> name When working with two features, : ID?<CR> please note that only the linking : MAHR GMBH,S1840 <CR><LF> formula for the first feature can be Requesting the software version number changed. : VER?<CR> : n.nn_<CR><LF>...
  • Page 73: M1240 Interface Protocol

    Switching into the basic status (RESET) : R<CR> : R<CR> Starting the master measurement : Z<CR> : Z<CR> The two point measurement is started only. It must be ended by pressing any key on the keyboard. Mahr GmbH, Millimar S1840...
  • Page 74 The password itself (the seven digits) :P91,Mn,-xxx.xxx<CR> can only be entered on the Millimar Setting the nominal value (see also section 11). :P91,Mn, xxx.xxx<CR> The password itself can neither be set :P91,Mn,xxx.xxx<CR> nor changed from the PC. Mahr GmbH, Millimar S1840...
  • Page 75: Optorss Interface Protocol

    : ?<CR> D 1000 S program. : xxx.xxx<Unit><CR><LF>, The data format matches that of the numerical measured value display. No data is sent when the catalog of functions and parameters is open or during master measurement. Mahr GmbH, Millimar S1840...
  • Page 76: Connecting To The Pc Using A Zero Modem Cable (Basic Circuit Diagram)

    14.7 Connecting to the PC using a zero modem cable (basic circuit diagram) Mahr GmbH, Millimar S1840...
  • Page 77: Working With The Parallel (I/O) Interface

    Recorders, devices with analog/digital converters OUT-2 opt. Output A2 (+) or control units with analog inputs can be OUT-2 opt. Output A2 (-) connected via the analog output. OUT-1 opt. Output A1 (+) OUT-1 opt. Output A1 (-) Mahr GmbH, Millimar S1840...
  • Page 78: Digital Control Output

    A1 = Signal "Valid" A2 = Signal "Rework" A3 = Signal "Reject" MODE 5 A1 = Signal "Measurement is valid" A2 = Signal "Measuring value < Lower warning limit" A3 = Signal "Measuring value > Upper warning limit" Mahr GmbH, Millimar S1840...
  • Page 79: Digital Control Input

    In case the Millimar function assignment differs from the standard modes, the MODE 4* Millimar transmits the "Mode 5" Enables the connection of the Millimar S 1840/ signal while uploading data to the SG control unit. D1000S. This means for the D1000S E1 = Measurement period** „Do not override the Millimar function...
  • Page 80 M1SEL Displays feature 1 (if "Autodetect" is on). M2SEL Displays feature 2 (if "Autodetect" is on). CLRSTO Deletes all values in the measuring value memories. SENDSTO Transmits all values out of the measuring value memories. Mahr GmbH, Millimar S1840...
  • Page 81: Application Examples For Using Digital Control Inputs And Outputs

    IN3 - IN1- IN2+ 25 pin Sub-D connector IN2 - Fig. 10 IN1+ Connecting the Millimar to a PLC IN1 - 25 pin Sub-D connecto r Fig. 11 Connecting lights for classification of the measuring results Mahr GmbH, Millimar S1840...
  • Page 82: Analog Output

    D/A converter, there is a certain delay between the input signal and the voltage output at the analog output. Cable 100 W Screening 25 pin Sub-D connector Fig. 12 Basic circuit diagram of an analog output of the parallel (I/O) interface Mahr GmbH, Millimar S1840...
  • Page 83: 15.4.1 Selecting The Feature To Be

    START. The current measuring value or result is displayed again. If further settings are required, use the ESC, MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 84: Analog Output

    START. If further settings are required, use the ESC, MASTER, DATA and MENU keys to navigate to the corresponding item in the catalog of functions and parameters and make the required settings. Mahr GmbH, Millimar S1840...
  • Page 85: Restoring Factory Settings

    (4.4. 1 )" is displayed. 3. Use the DATA and MASTER keys to set the unit of measurement and accept it by pressing START twice. The message "MAHR" is displayed and then the current measuring value. Mahr GmbH, Millimar S1840...
  • Page 86: Error Messages

    (3. 1 . 1 .2)" was selected under "NORMAL (3. 1 . 1 . 1 )" mode. PROCESS --> MEASURE However, the probe raw value with --> MEASMOD (3.1.1) for TEST is OK. recording the measuring values. Mahr GmbH, Millimar S1840...
  • Page 87: Technical Data

    0.2 µm or 0.3 % of the probe value displayed Switch repeatability limit for digital outputs 0.2 µm Hysteresis of digital outputs 0. 1 µm Temperature coefficient 0.005 %/°C Max. number of probes for connection S 1840 M, S 1840 F S 1840 PE Mahr GmbH, Millimar S1840...
  • Page 88 9 V, max. 100 mA Input pressure (only S 1840 PE) S 1840 PE 2 bar S 1840 PE Fed 2. 1 bar Regulated by a fine pressure regulator. Use only pure compressed air that is free of oil! Mahr GmbH, Millimar S1840...
  • Page 89: Contacts At Mahr

    Reutlinger Straße 48, 73728 Esslingen, Germany The duration and conditions of the warranty are set out in the general terms and conditions of delivery E-mail: mahr.es@mahr.de of Mahr GmbH or in the contract of sale. Telephone: (+49) 0711/9312-600 Fax: (+49) 0711/9312-725...
  • Page 90: Index

    Connecting to a PC ............75 Analog output .............. 81, 87 Connection formula ............34 Technical data ..............86 Contacts at Mahr ..............88 ANALOG (output) ............. 82 Controlling measuring duration using a time A-OUTP (4.7.3) ..............82 preset ..................63 ASCII (interface protocol) ..........
  • Page 91 ESC (key) ................... 14 Data transfer to computer ........70 Excitation voltage ............... 86 Data transfer to printer ..........70 Interface protocol ............... 66 Interface record ..............64 I/O (4.7) ..................82 I/O interface ................76 Mahr GmbH, Millimar S1840...
  • Page 92 NONE (data flow control) ..........68 NORMAL (measuring process) ....... 54 NORMAL mode ..............56 NORMAL (value to be displayed) ......36 Number of features ............22 Number of probes for connection ......86 Numerical values for parameter settings ..10 Mahr GmbH, Millimar S1840...
  • Page 93 RESOL. (4. 1 .2) ..........21, 24, 25, 26 Resolution ............... 11, 86 Resolution of the numerical display ..... 21 Restoring factory settings ..........84 RS 232 interface ..............17 RTS/CTS (data flow control) ........68 Run-out measurement ........... 61 Mahr GmbH, Millimar S1840...
  • Page 94 XON/XOF (data flow control) ........68 Static measurements ............11 Storage temperatures ..........2, 88 Sum measurement ............60 Supply voltage ..............3, 86 Zero modem cable............75 Switch repeatability limit ..........86 Symbols in the instructions ..........1 Mahr GmbH, Millimar S1840...
  • Page 95: Navigating In The Catalog Of Functions And Parameters

    (3) The function "MASTER" is only displayed if the "MAST 1P" setting was selected under PROCESS -> MASTER. (4) The functions "MSTR.MAX" and "MSTR.MIN" are only displayed if the "MAST 2P" setting was selected under PROCESS -> MASTER. Mahr GmbH, Millimar S1840...
  • Page 96 PLAUS. + numerical value 1.6.1 PLAUS. - numerical value 1.6.2 DET LIM DETLIM+ numerical value 1.7.1 DETLIM- numerical value 1.7.2 MASTER numerical value MSTR MA numerical value MSTR MI numerical value see FEATURE/ FEAT. 2 FEAT. 1 Mahr GmbH, Millimar S1840...
  • Page 97 HIGH 4.1.3 4.1.3.1 MIDDLE 4.1.3.2 4.1.3.3 RANGE 1 |-----| 4.1.4 4.1.4.1 +/-10 4.1.4.2 +/-3 4.1.4.3 +/-1 4.1.4.4 +/-0.3 4.1.4.5 +/-0.1 4.1.4.6 +/-0.03 4.1.4.7 +/-0.01 4.1.4.8 +/-0.003 4.1.4.9 AUTO_R 4.1.4.0 ORIGIN1 CENTER 4.1.5 4.1.5.1 BOTTOM 4.1.5.2 4.1.5.3 Mahr GmbH, Millimar S1840...
  • Page 98 4.6.1.4 OPTORSD 4.6.1.5 MARTALK 4.6.1.6 FORMAT 8-N-1 4.6.2 4.6.2.1 7-O-2 4.6.2.2 7-E-2 4.6.2.3 HANDSHK NONE 4.6.3 4.6.3.1 XON/XOF 4.6.3.2 RTS/CTS 4.6.3.3 BAUD 38400 4.6.4 4.6.4.1 19200 4.6.4.2 9600 4.6.4.3 4800 4.6.4.4 2400 4.6.4.5 1200 4.6.4.6 4.6.4.7 Mahr GmbH, Millimar S1840...
  • Page 99 FACTOR numerical value 4.7.3.2 CLR-SEL CLEAR ? LOESCHEN 4.8.1 ZERO LO NEG VAL POS VAL CHAN 1 CALIBR 4.9.1 MIN. MSTR MAX. MSTR PASSWRD OFFS LO ADJ LO CHAN 2 1000000 4.9.1 (current value) (current value) Mahr GmbH, Millimar S1840...
  • Page 100 Mahr GmbH, Millimar S1840...
  • Page 101 Mahr GmbH, Millimar S1840...