Frequency
Max. measurement range
Update time
Minimum detectable distance
Influence of ambient temperature
Dielectric constant
Memory:
• non-volatile EEPROM
• no battery required.

Interface

• HART
• configuration:
• analog output:
• display (local):

Programmer (infrared keypad)

Siemens Milltronics Infrared IS (Intrinsically Safe) Handheld Programmer for hazardous
and all other locations (battery is non-replaceable)
• approval:
• ambient temperature:
• interface:
• power:
• weight:
• color:
1.
Referenced from the sensor reference point.
2.
See 'Near Blanking' on page 93 for more details.
Page 8
SITRANS Probe LR – INSTRUCTION MANUAL
5.8 GHz (6.3 GHz in N. America): refer to product
nameplate for confirmation
1)
20 m (65 ft)
mA output and loop display is updated once per
second
1)2)
0.3 m (1 ft), plus the shield length (if any)
0.003% / K
ε
> 3 (for < 3 use waveguide antenna or stillpipe)
r
Siemens SIMATIC PDM (PC), or
HART handheld communicator, or
Siemens infrared handheld programmer
4 to 20 mA, ± 0.02 mA accuracy
multi-segment alphanumeric liquid crystal with bar graph
(representing level)
ATEX II 1 G, EEx ia IIC T4, certificate SIRA 01ATEX2147
FM/CSA Class 1, Div. 1, Groups A, B, C, D
−20 to +40 °C (−5 to +104 °F)
proprietary infrared pulse signal
3 V lithium battery
150 g (0.3 lb)
black
7ML19985HR02