Changing a Parameter Value

Notes:
• Security must be disabled to enable programming: set P000 to the Unlocked Value
stored in P069.
• Invalid entries will be rejected or limited.
• CLEAR
can be used to clear the field.
1.
Use the ARROW keys
DISPLAY
2.
Key in the new value.
3.
Press ENTER
Parameter Reset to Factory Default
1.
Scroll to the parameter or enter its address.
2.
Press CLEAR
P999 Master Reset
Following a Master Reset, complete reprogramming is required.
Note:
Resets all parameters to their factory settings, with the following exceptions:
P000 and P069 are not reset.
P838 is not reset (the learned TVT curve is not lost).
1.
Press PROGRAM
2.
Press DISPLAY
3.
Key in 999.
Press CLEAR
initiate reset. The LCD displays C.ALL.
4.
Reset complete. (Reset takes several seconds
to complete.)
7ML19985HR02
to scroll to the parameter number, or press
and key in the parameter number followed by ENTER
to set the value.
then ENTER
. The value returns to the default setting.
then DISPLAY
,
to open parameter fields.
then ENTER
, to Clear All and
SITRANS Probe LR – INSTRUCTION MANUAL
to activate PROGRAM mode.
.
Page 23