 This function is only available in continuous rotation.
 This function is not activated during reverse rotation.
When the mark of remaining amount of battery indicates "
supplied by the motor handpiece load may not be enough to reach the preset torque limit
value. In this case, the auto reverse function will not be activated. Should high torque be
required, use the AC adapter or use the device when the mark of the remaining amount of
battery indicates "
 If a load is continuously applied to the motor handpiece, it may automatically stop to
prevent overheating. In this case, leave the motor handpiece for a while until it cools down.
6.7.

Selecting a File System

To choose a different file system, press the SYSTEM  or  key. The file system shown in
the display is the selected system.

6.7.1. Continuous Rotary File Systems

When a file system has been selected, the first file of the system will automatically be shown in
the display.
Press the FILE  key to select the next file.
Press the FILE  key to select the previous file.
WARNING
Do not use files designed for reciprocating motion in continuous rotation.
FISDR-JAP / F1902124.EN / 07 / 2011 – updated 10/2016
").