Pay attention to the direction of the battery connector when installing. Forcible setting in the
wrong direction may cause damage and fluid leakage due to a short circuit.
Fully-charged rechargeable batteries generally discharge gradually over time even though
the device is not used. It is recommended to recharge the battery just before use.
Should the device stop automatically due to low battery voltage, it may not indicate low
voltage immediately when the device is powered on again.
Recharge rechargeable battery after it depletes as much as possible. Repeating short-time
use and subsequent recharging may shorten its operating time due to the "memory effect."
The battery may recover after repeating complete discharge and full charge for a few times
(see chapter 6.9. Battery Refresh).
The used Nickel Metal Hydride batteries are recyclable, but their disposal may sometimes
not be permitted by law in your location. Return them to your distributor.
When disposing of the control unit, follow the instructions of your local government for
disposal, as they contain materials which may become industrial waste.
When discarding the contra-angle and motor handpiece, dispose of them as medical
waste.
This product does not consider a patient's age, gender, weight or nationality.
No special training is required for this device.
The manufacturer declines any responsibility in the case of:
Use of the device for applications other than those specified in the instructions for use and
maintenance.
Modifications or repairs performed by persons not authorized by the manufacturer.
Use of non-original components or components other than those specified in the
STANDARD COMPONENTS chapter (see chapter 6.1).
File breakage due to misuse.
Accessories or device breakages due to sterilization: none of the X-SMART Plus
components are sterilizable (except for the contra-angle).
5.
Adverse Reactions
There are no known adverse reactions.