Per FCC Part 15.21, changes or modifications not
expressly approved by the party responsible for
compliance could void the user's authority to operate
this equipment.
Failure to follow recommendations for environmental
operating conditions, including input water temperature,
could result in injury to patients or users.
Insufficient water flow could result in elevated water
and tip temperature. When operated at the input water
temperature specified in the Water Line Requirements
Section 7.1 and with sufficient water flow, the water and
tip temperature should not exceed 50
SECTiON 4: Precautions

4 1 System Precautions

Do not place the system on or next to a radiator or
other heat source. Excessive heat may damage the
system's electronics. Place the system where air is free
to circulate on all sides and beneath it.
The system is portable, but must be handled with care
when moving.
Equipment flushing and dental water supply system
maintenance are strongly recommended. See Section
10: System Care.
Close manual shut-off valve on the dental office water
supply every night before leaving the office.
The use of an in-line water filter is recommended.
The use of an air dryer on the compressor line supplying
the System will prevent condensation from forming in
the air line which in turn may cause "caking" of the air
polishing powder and clogging of the lines and air
polishing nozzle.
Cavitron
Prophy Powders are specially formulated for
®
use in Cavitron
Air Polishing Systems. Do not use any
®
other materials in the air polishing powder reservoir.
Empty the air polishing powder bowl at the end of the
day to prevent "caking" of the powder and clogging of
the lines and air polishing nozzle.
Residual prophy powder in threads of the bowl and cap
can result in excessive wear and disengagement of
the cap during unit operation. Be sure to clean threads
regularly as per Section 10 System Care.

4 2 Procedural Precautions

General
As with all dental procedures, use universal precautions
(i.e., wear face mask, eyewear, or face shield, gloves and
protective gown).

Ultrasonics

o
o
C (122
F).

Air Polishing

7
The Cavitron
JET Plus unit works with Cavitron inserts
®
as a system, and was designed and tested to deliver
maximum performance for all currently available Cavitron
brand ultrasonic inserts. Companies that manufacture,
repair or modify inserts carry the sole responsibility for
proving the efficacy and performance of their products
when used as a part of this system. Users are cautioned
to understand the operating limits of their insert before
using in a clinical setting.
Like bristles of a toothbrush, ultrasonic inserts "wear"
with use. Inserts with just 2 mm of wear lose about 50%
of their scaling efficiency. In general it is recommended
that ultrasonic inserts be discarded and replaced after
one year of use to maintain optimal efficiency and
avoid breakage. A DENTSPLY Professional Insert
Efficiency Indicator is enclosed for your use.
If excessive wear is noted, or the insert has been
bent, reshaped or otherwise damaged, discard the insert
immediately.
Ultrasonic insert tips that have been bent, damaged, or
reshaped are susceptible to in-use breakage and should
be discarded and replaced immediately.
Retract the lips, cheeks and tongue to prevent contact
with the insert tip whenever it is placed in the patient's
mouth.
Patients should wear safety glasses or eye protection
during air polishing treatment.
Patients wearing contact lenses should remove them
prior to air polishing treatment.
Patients who have severe respiratory illness should
consult their physician before undergoing air polishing
prophylaxis procedures.
Avoid use on cementum or dentin.
Direct contact of prophy powder with surfaces and
marginal areas of dental restorations should be avoided.
Set the air polishing powder flow control to the maximum
(H) position only when it is necessary to remove
particularly difficult stains. Return the powder flow
control to the medium position upon the completion of
the procedure.
JET Air Polishing Insert nozzles that have been bent,
damaged or re-shaped, are susceptible to in-use
breakage and should be discarded and replaced
immediately.
Check o-ring and threads on powder bowl cap to
ensure a tight seal. If o-ring or threads are worn,
replace immediately.