6.5.7. Sound Volume Adjustment
The sound volume can be adjusted to 3 different levels: high, low and limited off (the sound
appears with a low level at the time of confirmation and error, but it does not sound during
reverse rotation, nor when reaching the torque limit value.
1. Press the sound volume (12) key (see Fig. 1).
2. The sound volume and the volume mark on the LCD panel change.
NOTES
The last sound volume is retained, even if the power is turned off.
If "reset to default parameters" is carried out,
6.6.
Operation
6.6.1. File Library
The device contains a file library with the following preset NiTi systems:
A. Continuous rotary systems
Gates
Proglider
PathFile
Protaper Next
Protaper Gold
Protaper
Program (individual programs)
B. Reciprocating systems
WaveOne
WaveOne
RECIPROC
The manufacturer reserves the right to update the file library and the systems contained in it.
WARNINGS
Follow the file manufacturer's instructions for use of endodontic files.
The file system shown on the display must always match the file in use. This is of the
utmost importance in order to avoid misusing reciprocating files and continuous rotary files.
Torque and speed values are subject to change by the file manufacturers without notice.
Therefore, the preset values in the library must be checked prior to use. Torque values
shown on the display are accurate and reliable only with X-SMART Plus 6:1 contra-angle
properly maintained and lubricated.
FISDR-JAP / F1902124.EN / 07 / 2011 – updated 10/2016
®
®
®
™
Universal
®
Gold
®
®
ALL (RECIPROC
®
the sound is set to the high level.
and RECIPROC
blue)
®
®
- 21 / 47 -