DENTSPLY X-Smart Plus User Manual page 7

Table of Contents

Advertisement

 Do not disassemble or alter the device; Dentsply Maillefer declines any responsibility in
case of alteration or modification of the device.
 Do not expose the X-SMART Plus unit, motor handpiece or AC adapter to any liquid.
 Do not drop the device.
 The X-SMART Plus requires special precautions as regards electromagnetic compatibility
(EMC) and must be installed and commissioned in strict conformity with the EMC
information provided in this instruction manual. Specifically, do not use the device close to
fluorescent lamps, radio transmitters and remote controls.
 Portable and mobile radio frequency (RF) communications equipment can affect the proper
functioning of X-SMART Plus.
 In order to avoid possible risks due to electromagnetic interference, do not use any
electrical medical device or electrical device of any other kind in close proximity to the X-
SMART Plus. The electromagnetic radiation emitted by the device is below the
recommended limits set forth in pertinent regulations in force (EN 60601-1-2:2007).
 Do not use the device in the presence of free oxygen or anesthetic substances or
flammable products.
 The device may possibly malfunction if used in the presence of an electromagnetic
interference wave. Do not install the X-SMART Plus in the vicinity of any device that emits
magnetic waves.
 The use of accessories, transducers and cables other than those specified, with the
exception of transducers and cables sold by the manufacturer of the X-SMART Plus as
replacement parts for internal components, may result in increased emissions or
decreased immunity of the X-SMART Plus.
 The X-SMART Plus should not be used adjacent to or stacked with other equipment. In the
instance that if adjacent or stacked use is necessary, the X-SMART Plus should be
observed to verify normal operation in the configuration in which it will be used.
 None of the X-SMART Plus components are delivered disinfected or sterilized: components
such as control unit, micromotor and micromotor cable need to be disinfected, the contra-
angle needs to be sterilized prior to first use and in between each patient!
 Never place the motor handpiece or any other device accessories in an autoclave unit or
ultrasonic tank.
 None of the components of the X-SMART Plus may be sterilized (contra-angle excluded,
see chapter 6.12. Cleaning, Disinfection and Sterilization).
 Do not immerse the device in ultrasonic cleaners.
 The plastic enclosure is not sealed, do not use any liquid or spray directly on the unit,
especially on the monitor or near the electrical sockets.
 Do not crimp the cords exiting the motor handpiece and AC adapter.
This device is equipped with an electronic circuit to limit file separation. Nevertheless, files
may still separate due to incorrectly set torque or speed, or the use of fatigued files.
 Follow the file manufacturer's instructions for use of the endodontic files.
 The file system shown on the display must always match the file in use. This is of the
utmost importance in order to avoid misusing reciprocating files and continuous rotary files.
 Do not use files designed for continuous rotation in reciprocating motion.
 Do not use files designed for reciprocating motion in continuous rotation.
FISDR-JAP / F1902124.EN / 07 / 2011 – updated 10/2016
- 7 / 47 -

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents