 Torque and speed values are subject to change by the file manufacturers without notice.
Therefore, the preset values in the library must be checked prior to use. Torque values
shown on the screen are accurate and reliable only with X-SMART Plus 6:1 contra-angles
properly maintained and lubricated.
 The accuracy of movement provided by the motor is guaranteed only if the original X-
SMART Plus 6:1 contra-angle is used, properly maintained and lubricated (for more details
see chapter 6.11.2. Lubricating the Contra-angle).
 Do not use any other contra-angle or other reduction rate other than the original one.
 Do not insert any file in the contra-angle during calibration.
 Calibrate every time the contra-angle is lubricated or replaced after sterilizing, or at least
once a week (see chapter 6.5.6. Calibration).
 When lubricating the contra-angle, carefully check that no lubricant penetrates the motor
handpiece.
 Do not lubricate the motor handpiece for any reason, as lubricant contamination of the
motor handpiece might damage it and may have a strong negative effect on its safe
operation.
 Never introduce any foreign objects into the motor handpiece shaft.
 The motor handpiece may overheat if excessive force is applied. If the motor handpiece
overheats too often or overheating persists, contact your distributor.
 Before starting the motor handpiece, verify the correctness of the motor settings.
 The AC adapter must be supplied at a voltage ranging between: 100 – 240 V (+/- 10%), 47-
63 Hz. Use only original parts.
 Should any anomalies arise during operation, suspend work and contact your distributor.
 Clinical judgment needs to be applied by the end user of the device.
4.

Precautions

Read these safety precautions thoroughly prior to use. These precautions allow you to use the
product safely, preventing harm to you and others.
It is of the utmost importance that this manual is preserved for future consultation. The manual
must accompany the system in all cases of sale or other transfer in order that the new owner
may reference the precautions and warnings.
Gloves and a rubber dam are mandatory during the use of X-SMART Plus.
Refer to the WARNINGS chapter (see chapter 3) to verify any special care to exercise before
starting to use the complete device.
 The appliance must only be used with the manufacturer's original accessories.
 Before changing the contra-angle or file, turn off the power of the unit. Changing with the
power kept on may cause unintended rotation by accidental touch of the ON/OFF key.
Always clean the shank of the file to be installed. Allowing dirt to enter the chuck could
cause loss of concentricity and deterioration of chucking force.