Step 6. Install the Agilent LC system

9 Add the Multimode calibrant to the calibrant bottle.
Never use aliphatic, aromatic or halogenated hydrocarbons in the CDS. These solvents
C A U T I O N
are not compatible with the O-ring in the CDS.
10 Turn on the Triple Quadrupole LC/MSrupole LC/MS front power switch to

Step 6. Install the Agilent LC system

Before you begin, check that you have these parts:
• HPLC grade methanol, or better
• HPLC grade water, or better
1 Install the Agilent LC system. Follow its installation documents. Other LCs
2 Connect the remote start cable from the Triple Quadrupole LC/MS to a
3 Install an Agilent G1369 LAN Interface card into the LC detector module if
Do not use any other LAN interface card besides the G1369 (e.g. the older Jet Direct
C A U T I O N
card J4100A).
34
start the pump-down of the Triple Quadrupole LC/MS.
The foreline pump(s) will become quieter within a few seconds.
are not supported.
Remote Start connector on the LC sampler, pump or other detector module
only.
required. If a detector module is not present, install the Agilent G1369 LAN
Interface card into the pump module. Before you install the LAN Interface
card, set switches 5 and 6 to the ON position. This will force the Agilent
LAN Interface Card to use its default IP address and subnet mask.
The default IP address is 192.168.254.11 and the default subnet mask is
255.255.255.0.
Agilent 6400 Series Triple Quad LC/MS System Installation Guide