Software Installation and Configuration
Step 9. Install Quantitative Analysis Reporting
Step 9. Install Quantitative Analysis Reporting
• Refer to the Offline Qualitative and Quantitative Analysis Installation
You must use the Run as administrator option to install the program. Failure to do so
C A U T I O N
can result in an incomplete installation, which can cause the system to run unreliably.
Step 10. Configure Excel for MassHunter (optional)
• Refer to the Offline Qualitative and Quantitative Analysis Installation
Step 11. Configure the instrument
If available, configure the device from the Device Configuration tool in the
Data Acquisition program. Go to Tools > Device Config.
Some HPLC instrument setups can be configured only through the Device
Configuration tool from the Data Acquisition program. The Capillary Pump is
an example.
In this step, you configure the Triple Quadrupole LC/MS and the HPLC
instrument.
1 Open the Agilent MassHunter Workstation folder on the desktop, and then
2 Double-click the Instrument Configuration tool.
3 If you are prompted to log in to OpenLAB, provide the user name and
42
Guide to install the Quantitative Analysis Reporting program.
Do not install any other MassHunter program until instructed to do so.
Guide to verify the installation of Excel Add-Ins for MassHunter.
double-click Acq Tools.
password for an OpenLAB administrator account.
You are prompted if OpenLAB Shared Services is installed on this
computer.
Agilent 6400 Series Triple Quad LC/MS System Installation Guide